IDEXX EquiView

®

 Digital Radiography System Operator’s Guide 

9

becomes limiting and does not improve image quality after a certain level.

 

TIP:

 Moving the X-ray source closer to the body part, which increases the effective kVp, may 

have a better impact on the image than simply continuing to increase mAs.

• 

Equine foot: 

The equine foot is another body part that produces a great deal of scatter. To 

produce good images of the navicular bone, similar techniques to the above may be helpful. 
A grid is always helpful for images where a grid can be easily used, in respect of maintaining 
the geometric requirements. In addition, lead shielding of unwanted parts of the image can 
be helpful.

To expose the plate:

1.  Place the digital X-ray plate into your X-ray equipment’s holder for film cassettes or in another 

suitable position. The plate’s active imaging surface must face the radiation source. 

2.  Select suitable settings on your X-ray equipment for exposure time and accelerating voltage.
 

WARNING:

 National regulations for dealing with X-ray radiation and the operation of X-ray 

devices applies. Do not exceed the applicable national standards for the radiation doses. 
Pay particular attention to the safety of the personnel in the vicinity of the X-ray equipment.

 

TIP:

 Use high kV and low exposure time. Dose increases proportionally with the kV, but to 

second order with time.

 

TIP:

 Using old generators that do not have full-wave rectification and have exposure times 

over 0.2 seconds is not ideal. If you have been using a good quality X-ray generator in the 
past, setting kVp at 80 and mAs at 1.8 is ideal for most exams.

 

TIP:

 Most X-ray generators have a milliamperage of 15 mA, so the time factor would be set to 

0.12 seconds: 15 mA x 0.12 s = 1.8 mAs. For large structures, 80 kVp and an mAs-product 
of 15 mA x 0.25 s = .75 mAs can be used. 

.  Ensure that the desired structures are located between the X-ray source and the digital X-ray 

plate. The plate must be aligned vertically to the direction of incidence of the X-ray beams. 
The film-focal distance (FFD) should be in the range of 55–60 cm (21.6"–2.6").

 

TIP: 

The X-ray generator should be placed closely to the region of interest.

 

TIP:

 Movement artifact is usually a combination of movement of the X-ray source, the horse 

and the plate itself.

4.  Press and hold the exposure switch halfway down until you hear the beep indicating that the 

plate is ready for an exposure. This prepares the generator and plate, and can take up to 4 
seconds. In addition to the beep, the light on the plate turns from green to red, and a blue 
progress bar appears at the bottom of the computer screen and moves halfway across.

Take Images screen (X-ray mode)

Image area 

Borderline

Содержание EquiView

Страница 1: ...IDEXX EquiView Digital Radiography System Operator s Guide...

Страница 2: ......

Страница 3: ...or industrial property rights covering this document or subject matter in this document The furnishing of this document does not give a license to these property rights except as expressly provided i...

Страница 4: ......

Страница 5: ...ing Images 12 Manipulating Images 14 Dynamic Views 14 Printing Images 17 Printer Settings 17 Preview Mode 17 Exporting Images and Creating CDs 18 Exporting Images 18 Creating a Patient CD 18 Viewing t...

Страница 6: ......

Страница 7: ...ng processing and archiving digital X ray images Computer The portable Microsoft Windows compatible computer runs the EquiView software which lets you view manage and archive the image files created f...

Страница 8: ...ribed in the Accessories section and if a hazard free combination with these devices components or assemblies cannot be deduced from the technical data you must consult the respective manufacturer to...

Страница 9: ...s a risk of damage Wait until the EquiView Digital Radiography System has warmed up to the ambient temperature and condensation has completely dried This may take several hours The initial installatio...

Страница 10: ...tion use the trigger cable to connect the switch box s detector socket to the computer 5 Turn the switch box to the on position 6 Connect the power cord to the computer and then to an appropriate elec...

Страница 11: ...tware indicated on the software disc 3 On the initial screen of the software there are two options The Take Images option lets you select or add a patient and then take X rays The View Images option l...

Страница 12: ...configured so that horse names are not displayed until a specified number of characters have been typed 2 When you locate the desired horse record click the name and then click the Continue button to...

Страница 13: ...of the breed to quickly access a particular breed in the list This is an optional field you can accept the default value Unknown 8 In the Age field enter the horse s age or use the up or down arrows...

Страница 14: ...t likely error that can occur is underexposure in which only a part of the dynamic range of the digital X ray plate is used In this case especially bony structures appear significantly less structured...

Страница 15: ...quipment TIP Use high kV and low exposure time Dose increases proportionally with the kV but to second order with time TIP Using old generators that do not have full wave rectification and have exposu...

Страница 16: ...n with that view You can skip image views if necessary or desired by selecting the check box next to that particular image Reshooting an Image To reshoot an image click the view name it turns blue the...

Страница 17: ...age taken while in the Take X Ray mode However you can display all of the images for the particular view when you are in the View Images mode Reviewing Images You can double click any image view name...

Страница 18: ...s display in their corresponding sections Search by Owner Click the owner s name in the Owners Found section This displays all horses for that owner in the Horses Found section To view the images of a...

Страница 19: ...navigation bar at the bottom to select individual views TIP At the lower left corner of each image is a toolbar with the following tools View single and multiple images Measure distances and angles S...

Страница 20: ...tings for the images and make the diagnosis Dynamic Views The mathematical treatment of the raw data from the plate is a very complex combination of functions known as an algorithm IDEXX has developed...

Страница 21: ...ossible osteophyte or enthesiophyte formation around the joint The dynamic range of the digital X ray plate and the EquiView software allow all these details to be seen in one image To adjust the Wind...

Страница 22: ...ld the Z key on the keyboard and then scroll with the mouse Or right click and scroll up Rotate To rotate an image counterclockwise press and hold the A key on the keyboard To rotate the image clockwi...

Страница 23: ...commend printing the images in Landscape as this allows the image to fill most of the paper size 2 Select the printer and choose the paper size 3 Click Print NOTE The EquiView Digital Radiography Syst...

Страница 24: ...ee types of images to indicate how you want the images to be exported 4 Select the location where you want to save the exported images NOTE The default location for exported images is the Images to Ex...

Страница 25: ...stalled on your computer If the software is not already installed follow the instructions below To install the Microsoft NET Framework software 1 The patient CD is equipped with the NET Framework soft...

Страница 26: ...ecting Disable Auto Protect You will be asked to choose a duration select Until system restart 3 Stop the SQL Server Service by right clicking the SQL Server icon and selecting MSSQLServer Stop When y...

Страница 27: ...r symbols Click the icon and click Properties The device should be enabled If the device is disabled click Enable If this does not solve the problem try restarting the computer by turning it off and w...

Страница 28: ...osive solvent or gaseous disinfectants If you use disinfectants that form an explosive gas mixture they must evaporate first before turning on the device again Turn off the power before disinfecting t...

Страница 29: ...m IDEXX Technical Support The equipment can only be used again after repair Operation with defective components can lead to an increased safety risk Interval Scope Method Daily Damaged parts inscripti...

Страница 30: ...ced into the case between taking shots A combined heating and cooling element Peltier element is installed in the transport case and insulation ensures that the temperature within the case changes ver...

Страница 31: ...te After connecting the power supply it is absolutely necessary that the temperature control of the case be switched on using the switch and the case closed This ensures that the valuable plate has a...

Страница 32: ...rtable unit Active area 20 cm x 25 cm 8 x 10 Electrical Data Computer Power supply 100V 240V AC 10 50 60 Hz Plate case Power supply 100V 240V AC 10 50 60 Hz 75 Wh power consumption or 12V DC using the...

Страница 33: ...nsation can form in the device Operate the device only under the above mentioned environmental conditions Computer Specifications CPU Intel Pentium 4 processor 2 4 GHz 512 KB cache 400 MHz system bus...

Страница 34: ...e your finger left to right and up and down on the trackpad to adjust the settings for window and level respectively Zoom Right click and hold and then move your finger up and down on the trackpad to...

Страница 35: ...1 1 5 20 24 DP 66 0 12 1 8 20 24 Carpus All views 66 0 1 1 5 20 24 Skyline 68 0 14 2 1 12 16 Tarsus Lat Obl 70 0 1 1 5 20 24 DP 70 0 12 1 8 20 24 Stifle Lat Obl 70 0 2 3 20 24 CC 70 0 4 6 close as po...

Страница 36: ...One IDEXX Drive Westbrook Maine 04092 USA 06 14899 00...

Отзывы: