Form Number TM61096 

Rev 2 Nov 2010 

Physical and Environmental Characteristics 

Characteristics 

Description 

Dimensions (W×H×L) 

84mm×188mm×41mm 

Weight (with battery) 

230g 

Environmental characteristics 

Description 

Temperature operating 

0 to +50°C 

Non-Operating 

-20 to +60°C 

Humidity (operating) 

<80% R.H. 

Altitude Operating 

2,000M (6560 ft.) 

Non-Operating 

12,300M (40354 ft.) 

Vibration & shock Operating 

MIL-T-28800E TYPE II Class 5 
2.66gRMS, 5 to 500 Hz, 3axes 
(10 minutes each) 

Indoor Use 

Indoor Use 

Certifications and compliances 

Safety 

IEC 61010 and designed to meet UL 3111 specifications 

Input rating 

V / Ω: Category 

III

 1000 Volts 

Over voltage category 

CAT 

III

: Distribution level mains, fixed installation. 

CAT 

II

: Local level mains, appliances, portable equipment 

CAT 

I

: Signal level, special equipment or parts of equipment, 

telecommunication, electronics. 

Pollution Degree 2 

Do not operate in environments where conductive 
Pollutants may be present. 

EC Declaration 
of Conformity 

Meets the intent of Directive 89/336/EEC for Electromagnetic 
Compatibility and Low Voltage Directive 73/23/EEC for specifica -
tions as listed in the official Journal of the European Communites: 
En 55011 Class A: Radiated and Conducted 

Emissions. 

En 50082-1 Immunity: IEC 801-2 Electrostatic Discharge 

IEC 801-3 RF Radiated 

En 61010-1 Safety requirements for electrical equipment for 

measurement, control, and laboratory use. 

Содержание SPLIT-JAW 61-096

Страница 1: ...mation Precautions and safety information Specifications Basic maintenance cleaning replacing the battery and fuses Performance test procedures Calibration and calibration adjustment procedures Form N...

Страница 2: ...ification 4 Current Specification 4 Physical and environment characteristics 5 Certification and compliance 5 Required Equipment 6 Basic Maintenance 7 Opening the Meter Case 7 Replacing the Battery 7...

Страница 3: ...mation Use the Meter only as described in the Service Manual If you do not do so the protection provided by the Meter may be impaired Read the Safety Information page before servicing this product In...

Страница 4: ...es that could result in personal injury or loss of life Specific precautions Do not operate without covers To avoid personal injury do not apply any voltage or current to the product without covers in...

Страница 5: ...ption Display count 6000 Numeric update rate 5 times sec Polarity display Automatic Overrange display OL is display Low voltage indicator is indicated Automatic power off time Automatic backslit off 3...

Страница 6: ...r input voltage up to 30V Impedance increase with input voltage to 330K at 1000V Resistance Continuity Function Range Accuracy 0 0 99 9 2 reading 1 100 2000 2 reading 2 digits Overload protection 750V...

Страница 7: ...ategory III 1000 Volts Over voltage category CAT III Distribution level mains fixed installation CAT II Local level mains appliances portable equipment CAT I Signal level special equipment or parts of...

Страница 8: ...should be performed only by qualified personnel Table B Required Equipment Equipment Required Characteristics Recommended Model Calibrator AC Voltage Range 0 750V AC Accuracy 0 07 Basic Frequency Rang...

Страница 9: ...pen circuit connections are exposed To open the Meter case and do the following 1 Disconnect test leads from any live source push the button to OFF and remove the test leads from the front terminals 2...

Страница 10: ...dicated If the Meter fails any of these tests it needs calibration adjustment or repair Testing the voltage Function To verify accuracy in the AC and DC voltage ranges do the following 1 Push the butt...

Страница 11: ...esistance Function To verify the accuracy of the resistance function do the following 1 Connect the calibrator to V and COM on the Meter 2 Push the button to start your Meter 3 Apply the inputs for st...

Страница 12: ...to start your meter 3 Apply the inputs for steps 1 4 in Table 4 4 For each input compare the readings on the Meter display to the reading in Table 4 5 If the display reading falls outside of the range...

Страница 13: ...potentiometer VR13 until display reads 1005 1006 3 Set the output of calibrator for resistance 100 and connect to V and COM test probes 4 Using a small flat tip screwdriver adjust the potentiometer V...

Страница 14: ...Form Number TM61096 Rev 2 Nov 2010 Figure 1 61 096 Calibration Adjustment Points 12 VR13 VR12 VR15 VR17 VR16 VR14...

Страница 15: ...Form Number TM61096 Rev 2 Nov 2010 Figure 2 61 096 Calibration Adjustment Points 13 VR1...

Отзывы: