Form Number TM61096
Rev 2 Nov 2010
Testing the AC Current Function
To verify the accuracy of AC current measurement functions, do the following:
1. Open jaw of the meter around the suitable wire or conductor.
2. Push the button to start your meter.
3. Apply the inputs for steps 1-4 in Table 4.
4. For each input, compare the readings on the Meter display to the reading in Table 4.
5. If the display reading falls outside of the range shown in the Table 4, the meter does not meet
specification.
Table 4 AC Current Test:
Step
Source
Frequency
Reading
1
100A
50Hz
97.9 to 102.1
2
100A
60Hz
97.9 to 102.1
3
200A
50Hz
196.1 to 203.9
4
200A
60Hz
196.1 to 203.9
10