Form Number TM61096
Rev 2 Nov 2010
Required Equipment
Required equipment is listed in Table B. If the recommended models are not available, equipment with
equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Table B. Required Equipment
Equipment
Required Characteristics
Recommended Model
Calibrator
AC Voltage Range:
0
~
750V AC
Accuracy:
±0.07% (Basic)
Frequency Range:
40
~
1KHz
Accuracy:
±2%
DC Voltage Range:
0
~
1000V DC
Accuracy:
±0.006% (Basic)
Current Range:
0
~
10A
Accuracy:
AC (40Hz to 1KHz): ±0.08% (Basic)
Resistance Range:
1Ω
~
100MΩ
Accuracy:
±0.03% (Basic)
1. Fluke 5700A Calibrator
2. 50 Turns Coil
6