Form Number TM61096 

Rev 2 Nov 2010 

CALIBRATION PROCEDURE 

Recalibrate your meter: 
It is recommended that the meter may be calibrated once year. Use the following procedure to calibrate the 
clamp meter. 

1. Perform calibration at an ambient temperature and relative humidity (23°C±2°C and R.H.≤80%). Allow

instrument to sit at this temperature for at least thirty minutes.

2. Disconnect the test leads from any circuit under test and turn off meter.
3. Loosen the screws from the battery cover, remove battery cover and pull the battery out of battery box

with battery still connected.

4. Loosen the two screws from the bottom case.
5. Lift the bottom case.

(A) DCV Calibration (Adjust VR15) 

1. Press Hold button of the meter to power on.
2. Set the output of calibrator for DCV 100V and connect to V and COM test probes.
3. Using a small flat-tip screwdriver adjust the potentiometer VR15 until display reads 100.1V

~

100.2V.

4. Disconnect the output terminals of DCV calibrator.

(B) Resistance Calibration (Adjust VR13 & VR14) 

1. Set the output of calibrator for resistance 1KΩ and connect to V and COM test probes.
2. Using a small flat-tip screwdriver adjust the potentiometer VR13 until display reads 1005Ω

~

1006Ω.

3. Set the output of calibrator for resistance 100Ω and connect to V and COM test probes.
4. Using a small flat-tip screwdriver adjust the potentiometer VR14 until buzzer sounds and check 400Ω

buzzer don’t sounds.

5. Disconnect the output terminals of resistance calibrator.

(C) ACV Calibration (Adjust VR16& VR1) 

1. Set the output of ACV calibrator for ACV 70V 60Hz and connect to V and COM test probes.
2. Using a small flat-tip screwdriver adjusts the potentiometer VR16 until display reads 70.0V±1d.
3. Set the output of ACV calibrator for ACV 200V 60Hz and connect to V and COM test probes.
4. Using a small flat-tip screwdriver adjusts the potentiometer VR1 until display reads 199.8V±1d.
5. Disconnect the ACV calibrator from the meter.

(D) ACA Calibration (Adjust VR17 & VR12) 

1. Flow the current of ACA 100.0A 60Hz around the suitable wire or conductor.
2. Using a small flat-tip screwdriver adjusts the potentiometer VR17 until display reads 99.8A.
3. Flow the current of ACA 300.0A 60Hz around the suitable wire or conductor.
4. Position the clamp around the wire or conductor, release clamp trigger to make sure that the clamp is

entirely closed.

5. Using a small flat-tip screwdriver adjusts the potentiometer VR12 until display reads 300.0A.
6. Remove the meter away from the wire or conductor.

11 

Содержание SPLIT-JAW 61-096

Страница 1: ...mation Precautions and safety information Specifications Basic maintenance cleaning replacing the battery and fuses Performance test procedures Calibration and calibration adjustment procedures Form N...

Страница 2: ...ification 4 Current Specification 4 Physical and environment characteristics 5 Certification and compliance 5 Required Equipment 6 Basic Maintenance 7 Opening the Meter Case 7 Replacing the Battery 7...

Страница 3: ...mation Use the Meter only as described in the Service Manual If you do not do so the protection provided by the Meter may be impaired Read the Safety Information page before servicing this product In...

Страница 4: ...es that could result in personal injury or loss of life Specific precautions Do not operate without covers To avoid personal injury do not apply any voltage or current to the product without covers in...

Страница 5: ...ption Display count 6000 Numeric update rate 5 times sec Polarity display Automatic Overrange display OL is display Low voltage indicator is indicated Automatic power off time Automatic backslit off 3...

Страница 6: ...r input voltage up to 30V Impedance increase with input voltage to 330K at 1000V Resistance Continuity Function Range Accuracy 0 0 99 9 2 reading 1 100 2000 2 reading 2 digits Overload protection 750V...

Страница 7: ...ategory III 1000 Volts Over voltage category CAT III Distribution level mains fixed installation CAT II Local level mains appliances portable equipment CAT I Signal level special equipment or parts of...

Страница 8: ...should be performed only by qualified personnel Table B Required Equipment Equipment Required Characteristics Recommended Model Calibrator AC Voltage Range 0 750V AC Accuracy 0 07 Basic Frequency Rang...

Страница 9: ...pen circuit connections are exposed To open the Meter case and do the following 1 Disconnect test leads from any live source push the button to OFF and remove the test leads from the front terminals 2...

Страница 10: ...dicated If the Meter fails any of these tests it needs calibration adjustment or repair Testing the voltage Function To verify accuracy in the AC and DC voltage ranges do the following 1 Push the butt...

Страница 11: ...esistance Function To verify the accuracy of the resistance function do the following 1 Connect the calibrator to V and COM on the Meter 2 Push the button to start your Meter 3 Apply the inputs for st...

Страница 12: ...to start your meter 3 Apply the inputs for steps 1 4 in Table 4 4 For each input compare the readings on the Meter display to the reading in Table 4 5 If the display reading falls outside of the range...

Страница 13: ...potentiometer VR13 until display reads 1005 1006 3 Set the output of calibrator for resistance 100 and connect to V and COM test probes 4 Using a small flat tip screwdriver adjust the potentiometer V...

Страница 14: ...Form Number TM61096 Rev 2 Nov 2010 Figure 1 61 096 Calibration Adjustment Points 12 VR13 VR12 VR15 VR17 VR16 VR14...

Страница 15: ...Form Number TM61096 Rev 2 Nov 2010 Figure 2 61 096 Calibration Adjustment Points 13 VR1...

Отзывы: