Specifications

1-1 General Specifications

Jaw opening:

  0.5” (12.7mm) Accepts up to 1/0 copper 

cable.

LCD display digits:

  6000 digit large scale LCD readout.

Measuring rate:

  5 times / sec.

Overrange display:

  “OL” is displayed for “

Ω

” function, 

shows the real value for “A” and “V” function.

Automatic power off time:

  Approximately 30 minutes 

after power on.

Low battery indicator:

          is displayed. Replace the 

battery when the indicator          appears in the display.

Power requirement:  

9V battery

Battery type and life:  

ALKALINE PP3, 9V 250 hours

1-2 Environmental Conditions

Indoor Use.

Calibration:  

One year calibration cycle.

Operating temperature:

  0°C ~ 30°C (<80% RH)

 

 

30°C ~ 40°C (<75% RH)

 

 

40°C ~ 50°C (<45%RH)

Storage temperature:  

-20 to +60°C, 0 to 80% RH 

(batteries not fitted)

Operating altitude:

 2000m (6562 ft)

Overvoltage category:

 

IEC 61010-1 CAT IV 600V

 

CAT III 1000V

IEC 61010-1 CAT IV 600V

CAT IV equipment of OVERVOLTAGE CATEGORY IV is for use at 

the origin of the installations. Note examples include electricity 

meters and primary over-current protection equipment.

Conductor Size:

  27mm diameter

Pollution degree:

  2

EMC:

  EN 61326-1

Shock vibration:

  Sinusoidal vibration per MIL-T-

28800E (5 ~ 55 Hz, 3g maximum)

Drop Protection:

 4 foot drop to hardwood on concrete floor

1 - 3 Electrical Specifications

Accuracy is ± (% r number of digits) at 23°C ± 5°C 

< 80%RH.

Temperature coefficient:

0.2 x (Specified accuracy) / °C, < 18°C, > 28°C.

Voltage

Function

Range

Accuracy

V~

1.3 ~ 750.0V

 ±(1.5%+3dgt)

50Hz ~ 500Hz

    V    

1.8V ~ 1000V

±(1% + 2 dgt)

-0.6V ~ -1000V

±(1% + 4 dgt)

Overload protection: 

1000 VDC / 750 VAC

Max Operation time:  

DT=30s for 

30V

Input impedance:

 

4K for input voltage up to 30V. 

Impedance increases with input voltage to approximately 

200K

Ω

 at 600V.

...

Resistance & Continuity

Function

Range

Accuracy

Ω

  

)

)

)

)

0.0

Ω

~99.9

Ω

±(2% +1

Ω

)

100

Ω

~2000

Ω

±(2% +2 dgt)*

*Temperature coefficient multiplies by 1.5 when operating 

temperature is 40°C ~ 50°C.

Overload protection:  

1000 VDC / 750 VAC

Max. open circuit voltage: 

1.5V

Continuity check: 

Internal beeper activates if the resistance 

of the circuit under test is less than 25

Ω

. It will then turn off 

if the resistance is increased beyond 400

Ω

.

AC Current

Function

Range

Accuracy

A~

0.6A ~ 200.0A

±(1.8% +3 dgt)*

50Hz ~ 60Hz

Overload protection:  

400A ±(1.8% +6 dgt) for <50A.

Max. open circuit voltage: 

1.5V

Continuity check: 

Internal beeper activates if the resistance 

of the circuit under test is less than 25

Ω

. It will then turn off 

if the resistance is increased beyond 400

Ω

.

Warranty Statement

This tester is warranted to the original purchaser against

defects in material and workmanship for two years from 

the date of purchase. During this warranty period, IDEAL

INDUSTRIES, INC. will, at its option, replace or repair the

defective unit, subject to verification of the defect or 

malfunction.

This warranty does not cover fuses, batteries or damage

from abuse, neglect, accident, unauthorized repair, alteration,

or unreasonable use of the instrument.

Any implied warranties arising out of the sale of an IDEAL

product, including but not limited to implied warranties of

merchantability and fitness for a particular purpose, are

limited to the above. The manufacturer shall not be liable

for loss of use of the instrument or other incidental or 

consequential damages, expenses, or economic loss, or for

any claim or claims for such damage, expenses or economic

loss.

State laws vary, so the above limitations or exclusions may

not apply to you. This warranty gives you specific legal

rights, and you may also have other rights which vary from

state to state.

Содержание SPLIT-JAW 61-096

Страница 1: ...attery cover Do not use without the battery and battery cover properly installed Do not attempt to repair this unit as it has no user serviceable parts Use the proper terminals functions and range for...

Страница 2: ...Off On HOLD HOLD HOLD AUTO SEL Press AUTO SEL HOLD HOLD HOLD Press 2sec COM V AUTOMATIC TESTER HOLD V AC 200A ACV 1 0V CAT IV 600V AUTO SEL AUTO SEL V AC Voltage AC DC Resistance Ohms DCV Measuring Ra...

Страница 3: ...formed by qualified personnel Cleaning Periodically wipe the case with a dry cloth and detergent Do not use abrasives or solvents Priority of Function 100A 200A CAT IV 600V 200A CAT IV 600V HOLD HOLD...

Страница 4: ...nput voltage to approximately 200K at 600V Resistance Continuity Function Range Accuracy 0 0 99 9 2 1 100 2000 2 2 dgt Temperature coefficient multiplies by 1 5 when operating temperature is 40 C 50 C...

Страница 5: ...a de la bater a No use el mult metro sin la bater a ni sin la tapa de la bater a correctamente instalada No intente reparar esta unidad ya que no tiene piezas reparables por el usuario Use las termina...

Страница 6: ...LD HOLD HOLD Press 2sec COM V AUTOMATIC TESTER HOLD V AC 200A ACV 1 0V CAT IV 600V AUTO SEL AUTO SEL V AC Apagado Apagado Encendido Encendido Pulse 30 minutos Pulse tiempo 2 segundos Voltaje de CA CC...

Страница 7: ...ioridad de las funciones 100A 200A CAT IV 600V 200A CAT IV 600V HOLD HOLD HOLD AUTOMATIC TESTER AUTOMATIC TESTER A AUTO SEL Press AC A AUTO SEL AC H A AUTO SEL AC H H A AUTO SEL AC La pantalla de cris...

Страница 8: ...e contaminaci n 2 Compatibilidad electromagn tica EMC Cumple con la norma EN 61326 1 Impacto y vibraci n Vibraci n sinusoidal de acuerdo con la norma MIL T 28800E 5 a 55 Hz 3g m ximo Protecci n contra...

Страница 9: ...r sent manuel d faut la protection offerte par le multim tre pourra tre compromise AVERTISSEMENT Pour viter tout risque d lectrocution de l sions personnelles ou de mort se conformer aux directives su...

Страница 10: ...e tension connue 2 Appliquer le multim tre au circuit en cours de contr le 3 Retourner la tension active connue pour v rifier le bon fonctionnement Travailler toujours avec un quipier Symboles figuran...

Страница 11: ...CAT IV 600V AUTO SEL AUTO SEL V AC 200A AUTO SEL HOLD CAT IV 600V COM V AUTOMATIC TESTER Plage de mesure 0 2000 Continuit Courant C A Amp res La m choire accepte les c bles cu jusqu au calibre 1 0 0 5...

Страница 12: ...jusqu 6000 Fr quence de mesure 5 fois s Affichage de d passement OL est affich pour la fonction la valeur r elle est affich e pour les fonctions A et V Coupure automatique de l alimentation Environ 30...

Страница 13: ...unit d fectueuse suite la v rification du d faut ou du dysfonctionnement Cette garantie ne s applique pas aux fusibles aux piles ou aux dommages r sultant d une utilisation abusive de la n gligence d...

Отзывы: