ABOUT CONTROL SOLUTION
Control solutions contain a known amount of glucose that reacts with test strips. By comparing your
control solution test results with the expected range printed on the test strip vial label, it is able to check
that the meter and the test strips are working together as a system and that you are performing the test
correctly. It is very important that you do this simple check routinely to make sure you get accurate results.
How often should the control solution test be performed?
•
When you use this system to test your blood for the first time, practice the procedure using
control solution. When you can do three tests in a row that are within the expected range, you are
ready to test your blood.
•
Routinely check the meter and test strips, so perform a single test for each level of control
solution at least once a week
When should the control solution test be performed?
•
When you first get your IDEAL LIFE Gluco-Manager™
•
When you begin using a new vial of test strips.
•
Whenever you suspect that the meter or test strips are not working properly.
•
When your blood glucose test results are not consistent with how you feel, or when you think your
results are inaccurate.
•
When your test strips are expeose to extreme environmental conditions (See “Important Test Strip
Information” below)
•
When you want to practice running the test.
•
If you drop the meter.
IMPORTANT CONTROL SOLUTION TEST INFORMATION
•
Use the same branded control solutions as the test strips
•
Check the expiry date on the control solution vial. Do not use if expired.
•
Control solution, meter, and test strips should come to room temperature
(20°C - 25°C / 68°F - 77°F) before testing.
•
Shake the vial before use. Discard the first drop of control solution. After squeezing, wipe off the
dispenser tip to avoid contamination. The above tips ensure you get a good sample and an
accurate result.
•
Use a period of 90 days from the date that you first open it. Record the discard date (date
opened plus 90 days) on the control solution vial. Discard after 90 days.
Installation
IDEAL LIFE Gluco-Manager™
18
Ideal Blood Gluco Manual 9/24/07 12:25 PM Page 18