WIKA operating instructions emission monitor for SF
6
gas, model GA35
14008241.01 01/2016 EN/DE
EN
6.9 Resetting to default settings
“DEFAULT” does not correspond to the factory settings (see table).
The factory settings can only be restored through the manual input of corresponding
values. The default settings will be saved permanently.
Basic setting
LOAD DEFAULTS
Factory setting upon deliv-
ery
Alarm High limit
95 %
50 % corresponds to 1,000 ppm
v
Alarm High idle polarity
LED ON
LED OFF
Alarm Low limit
5 %
25 % corresponds to 500 ppm
v
Alarm Low idle polarity
LED ON
LED OFF
LCD contrast
14
14
Resolution
5
5
1.
Select “SETUP”.
2.
Select “LOAD DEFAULTS”.
»
Instrument is reset to default settings.
6. System settings