WIKA operating instructions emission monitor for SF
6
gas, model GA35
14008241.01 01/2016 EN/DE
EN
5.6 Switching on
Requirement: Case completely closed.
▶
Connect the power supply.
»
Self test is carried out. For procedure, see table.
Time
Display
Alarm LED
Warning
tone
Alarm relay Error
relay
Analogue
output
0 ... 10 s
Empty
Off
On
Off
Off
Undefined
10 ... 20 s
Starts
Depending on
alarm setting
and gas
concentration
Off
Depending on
alarm setting
and gas
concentration
On
Approximate-
ly corre-
sponds to the
gas concen-
tration
> 20 s
Gas concen-
tration in
ppm
v
Depending on
alarm setting
and gas
concentration
Off
Depending on
alarm setting
and gas
concentration
On
Corresponds
to the gas
concentration
6. System settings
Personnel:
operating personnel
6.1 Menu navigation
Press the navigation key under the display to navigate through the menu.
The function of the navigation key will be shown in the navigation bar of the display.
The functions will vary depending on the context.
→ For the placement of the keys, see chapter 2.1 “Overview”
Navigation bar, context-dependent
function of the navigation key
5. Commissioning / 6. System settings