background image

Instructions Manual

9

================================
Program:

WRAPPED

Temperature: 134C
Pressure: 206.0 kPa
Drying Time: 08Min
Holding Time: 4.0Min
--------------------------------

Time

Temp.

Pressure

Start 12:28:17 089.0C
T1:

12:31:32 087.1C -075.0kPa

T2:

12:33:43 110.2C 052.0kPa

T3:

12:36:37 088.9C -075.0kPa

T4:

12:39:20 114.7C 053.7kPa

T5:

12:43:37 087.9C -075.0kPa

T6:

12:50:40 134.8C 206.0kPa

TS:

134.7C 209.5kPa

Max. Temperature:135.2C
Min. Temperature:134.3C
Max. Pressure:214.0kPa
Min. Pressure:204.9kPa
T7:

12:54:39 134.4C 211.4kPa

T8:

12:57:36 102.1C -060.0kPa

T9:

12:59:54 098.2C -060.0kPa

End

13:04:07 102.4C

--------------------------------
Cycle No.: 00017
Ster. Value: Success
Date: 2017-06-07
SN:E54723
Operator:
v 2B00V2.5
================================

=======================
Program:Vacuum test
Tp:1

P1:-75.0kPa
P2:-74.0kPa
rate of pressure rise:0.10
Start Time:08:22
End Time:09:01
Date:2017-07-19
Test Value:Success
SN:E00001
Operator:
========================

Содержание STE-23-D

Страница 1: ...STEAM STERILIZERS Instructions Manual For Model STE 23 D...

Страница 2: ......

Страница 3: ...ge in the manual Need Maintenance If this picture appears on the screen when power on or E88 appears on the report please call your dealer or local service maintenance Your steam sterilizer needs a re...

Страница 4: ...terilization program 6 5 3 Start the sterilization program 6 5 4 End of cycle 6 5 5 Manual abort the program 7 5 6 Test Programs 7 5 7 Record of the cycle 7 5 8 Printer 8 5 9 Report 8 6 Advance settin...

Страница 5: ...he maximum weight limit of the loads specified in this manual Do not operate this appliance if it has a damaged cord or plug if it is not working properly or if it has been damaged or dropped Never mu...

Страница 6: ...2 Main power switch 19 Condenser vent 6 Door spindle 13 Drain connector Distilled water tank 20 Bacteriological filter 7 Distilled water tank 14 Drain connector Used water tank 21 Rating plate 2 2 Con...

Страница 7: ...at level Max Approx 0 5 L Water at level Min Operation temperature 5 C 40 C Operation relative Humidity Max 80 non condensing Overall dimensions 490mm W 455mm H 690mm D Net weight 53 kg Max Noise lev...

Страница 8: ...supply connection Check the label on back panel o sterilizer to verify voltage rating for the unit Failure to connect the autoclave to an appropriate power supply could result in damage to the unit an...

Страница 9: ...esidual chemicals left over after cleaning and disinfecting process may damage and corrode parts of the autoclave always rinse off the instruments using distilled water Follow instrument manufacturer...

Страница 10: ...program the materials to be sterilized can now be placed on the tray placed inside the chamber by the tray handle After the instruments are loaded you may close the door 5 3 Start the sterilization pr...

Страница 11: ...he Helix test into the chamber then close the door and Tap button After finish the cycle you may check the indicator and evaluate the result according with the instructions of the test manufacture Vac...

Страница 12: ...ou can read the latest information of the last 20 cycles stored in the internal memory of the sterilizer 5 9 1 Select Report from the main menu and Tap button you will see the list of records 5 9 2 Se...

Страница 13: ...0kPa T6 12 50 40 134 8C 206 0kPa TS 134 7C 209 5kPa Max Temperature 135 2C Min Temperature 134 3C Max Pressure 214 0kPa Min Pressure 204 9kPa T7 12 54 39 134 4C 211 4kPa T8 12 57 36 102 1C 060 0kPa T...

Страница 14: ...old for 5 seconds This will enter into the advanced setting mode 6 1 Parameter Tap button to select Parameter Tap button to enter the menu Select the program that you need to change by Taping button T...

Страница 15: ...se it with distilled water and put it back in to the position 7 2 Replacement of the bacteriological filter The bacteriological filter is in the back of the sterilizer Unscrew the filter by hand anti...

Страница 16: ...r seal ring carefully with a smooth cloth with distilled water Moisten the new seal with distilled water Insert the new seal and Tap in sequence as follows 1 Press in the top and bottom of the door se...

Страница 17: ...le Contact your supplier if error persists E12 Failure to preheat the chamber Power off run a new cycle Contact your supplier if error persists E13 Vacuum failed Power off run a new cycle Contact your...

Страница 18: ...re Action Release of the steam and restoration of the safety pressure 4 Safety micro switch for the door status Comparison for the correct closing position of the door Action Signal of the wrong posit...

Страница 19: ...0mg kg Silicium oxide sio2 1mg I 1 0mg kg Iron 0 2mg I 0 1mg kg Cadmiun 0 005mg I 0 05mg kg Lead 0 05mg I 0 1mg kg Rest of heavy metals 0 1mg I 0 1mg kg Chloride 2mg I 0 1mg kg Phosphates 0 5 mg I 0 1...

Страница 20: ...id material 5 00 1 50 121 110 20 35 65 Single wrapped solid or hollow material 4 00 1 20 TEXTILE 134 210 8 45 65 Unwrapped porous material 1 25 0 40 Single wrapped porous material 1 10 0 30 121 110 30...

Страница 21: ...anual 17 Pressure kPa Solid Time min Pressure kPa Wrapped Textile Prion Time min Pressure kPa Time min Liquid Pressure kPa Time min B D Test HELIX Test Pressure kPa Vacuum Test Time min Pressure kPa D...

Страница 22: ...2 Air filter valve V3 Water pump valve V4 Water release valve V5 Vacuum pump start valve V6 Vacuum pump fan Chamber heater Steam heater Steam heater Chamber heater Thermal protector Water pump Vacuum...

Страница 23: ...iagram V1 Vacuum pump valve V2 Air filter valve V3 Water pump valve V4 Water release valve V5 Vacuum pump start valve Distilled water tank Air filter Used water tank Safety valve Chamber Test Port Con...