Instructions Manual
4
3. Installation
3.1 General conditions
Position the device on a plane surface with minimum capacity 60 kgs.
The sterilizer should be placed on a level worktable.
Improper water level in the chamber could cause a sterilizer malfunction.
Leave at least 10cm between the device rear part and the wall. The clearance required to open the door is 40cm.
Position the autoclave at such a height as to make it possible for the operator to check the whole sterilization chamber and
carry out the normal cleaning operations.
The room where the device is installed must be enough ventilated.
Do not install the device near washing basins, taps, etc. where it is likely to be splashed.
Do not lean on the door when it is opened.
Do not place trays , papers, fluid containers, etc. on the sterilizer.
3. 2 Power supply connection
Check the label on back panel o sterilizer to verify voltage rating for the unit. Failure to connect the autoclave to an
appropriate power supply could result in damage to the unit, and electrical shock to personnel.
Plug power cord into a properly polarized and grounded receptacle rated. A dedicated circuit only used for the sterilizer is
recommended.
Never connect the device pin to reductions of any type.
4 Setup
Open the door and remove all of the inner contents for unpacking.
Connect the power cord to an outlet of the appropriate voltage.
Turn on the main power switch on the right side. After switching on, the machine turns on the LCD and shows the door
position, water level, working program, date, time and etc.
Note: The control panel will be locked for the initial 10 seconds after powering up for system initialization.
Notice: Before using the sterilizer or at any time the low water level icon
blinks, fill the distilled water tank with
distilled water.
Allow Clearance
On Both Sides