![ICANCLAVE STE-23-D Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/icanclave/ste-23-d/ste-23-d_instruction-manual_3324123020.webp)
Instructions Manual
16
Appendix 2
Diagrams of the sterilization programs
The time required for sterilizer to be ready for routine use after the power is switched is less than 5
minutes.
The max. Temperature of the 134°C sterilization cycle is 137°C
The max. Temperature of the 121°C sterilization cycle is 124°C
Programs
Te
m
pe
ra
tu
re
(°
C)
Pr
es
su
re
(k
Pa
)
Ho
ld
in
g
tim
e
(m
in
)
To
ta
lt
im
e
(m
in
)
Type
M
ax
lo
ad
(k
g)
M
ax
lo
ad
pe
rt
ra
y
(k
g)
SOLID
134
210
4
25-45
Unwrapped solid material
5.00
1.50
121
110
20
30-60
WRAPPED
134
210
4
35-60
Unwrapped solid material
5.00
1.50
121
110
20
35-65
Single-wrapped solid or hollow
material
4.00
1.20
TEXTILE
134
210
8
45-65
Unwrapped porous material
1.25
0.40
Single-wrapped porous material
1.10
0.30
121
110
30
50-75
Dual-wrapped porous material
0.75
0.25
Single-wrapped hollow material
4.00
1.25
Dual-wrapped solid and hollow
material
2.00
0.60
PRION
134
210
18
50-75
Unwrapped porous material
1.25
0.40
Single-wrapped porous material
1.10
0.30
Dual-wrapped porous material
0.75
0.25
Single-wrapped hollow material
4.00
1.25
Dual-wrapped solid and hollow
material
2.00
0.60
LIQUID
(optional)
134
210
10
35-60
Liquid
1.20
0.40
121
110
30
35-65
Drying
(optional)
―
―
―
1-20
―
―
―
B&D test
134
210
3.5
28-35
―
―
―
Helix test
134
210
3.5
28-35
―
―
―
Vacuum
test
―
―
―
20-25
―
―
―
Содержание STE-23-D
Страница 1: ...STEAM STERILIZERS Instructions Manual For Model STE 23 D...
Страница 2: ......