Instructions Manual
11
7.Maintenance
To assure proper operation and maximum autoclave life, carefully follow all recommendations for periodic maintenance.
One of the MOST important steps you can take to prevent problems with your sterilizer is to use ONLY distilled water.
Frequency
Number of cycles
Maintenance operation
Monthly
50
Clean the door seal
Clean the filter inside the chamber and in the clean
water tank
Clean the chamber the trays and the rack
Clean the external surface
Every 3 months
200
Clean the distilled water tank
Every year
800
Replace the door seal
7.1 Clean the distilled water tank
Disconnect the main cable.
Drain the tank completely using the drain tube and leave it connected into the connector in a
open position.
Clean the internal surface with a soft sponge and a small soft brush for the areas that are
difficult to reach using and a mild soap.
Remove the filter and clean it with a small soft brush and mild soap, rinse it with distilled
water and put it back in to the position.
7.2 Replacement of the bacteriological filter.
The bacteriological filter is in the back of the sterilizer. Unscrew
the filter by hand anti-clockwise.
Place the new bacteriological filter.
Screw the new filter by hand clockwise.
Note: Do not operate sterilizer without filters in place.
7.3 Clean Chamber, door seal, trays and tray Rack.
Remove the trays and tray rack from the chamber.
Clean trays, rack and inside of chamber with mild soap.
Rinse the trays, rack and inside of chamber with a smooth cloth and
distilled water.
Examine door seal for possible damage.
Clean door seal and mating surfaces with a damp cloth.
Note: Do not use bleaching agents or any abrasive materials / substances
in chamber. Failure to comply may result in damage to the chamber and/or other components.
Caution: To prevent burns, let unit to cool before cleaning gaskets and touch the surface.
7.4 Door adjustment
Under normal circumstances the chamber door does not require adjustments. However, if
the seal fails (resulting in steam leaking from the front of the chamber), you may use the
spanner tool to tighten the door seal.