6-29
Chapter 6, COMPUTER OPERATION
byte is accessed by performing a serial poll.
Bit 0
Command complete
Bit 1
Invalid command
Bit 2
Command parameter error
Bit 3
Reference unlock
Bit 4
Overload
Bit 5
New ADC values available after external trigger
Bit 6
Asserted SRQ
Bit 7
Data available
N
Report overload byte
Causes the lock-in amplifier to respond with the overload byte.
Bit 0
not used
Bit 1
CH1 output overload (> ±120 %FS)
Bit 2
CH2 output overload (> ±120 %FS)
Bit 3
Y output overload (> ±300 %FS)
Bit 4
X output overload (> ±300 %FS)
Bit 5
not used
Bit 6
input overload
Bit 7
reference unlock
MSK [n]
Set/read service request mask byte
The value of n sets the SRQ mask byte in the range 0 to 255
REMOTE [n]
Remote only (front panel lock-out) control
Allowed values of n are 0 and 1. When n is equal to 1, the lock-in amplifier enters
remote only mode in which the front panel control functions are inoperative and the
instrument can only be controlled with the RS232 or the GPIB interfaces. When n is
equal to 0, the front panel control functions normally.
6.4.11 Instrument Identification
ID
Identification
Causes the lock-in amplifier to respond with the number 7220.
REV
Report firmware revision
Causes the lock-in amplifier to respond with the firmware revision number. This gives
a four line response which the controlling program must be able to accept.
VER
Report firmware version
Causes the lock-in amplifier to respond with the firmware version number. The
firmware version number is the number displayed on the front-panel RS232 SETUP 3
screen.
Содержание 7220
Страница 8: ...v i TABLE OF CONTENTS...
Страница 12: ...1 4 Chapter 1 INTRODUCTION...
Страница 76: ...5 36 Chapter 5 FRONT PANEL OPERATION...
Страница 124: ...D 2 Appendix D CABLE DIAGRAMS Figure D 2 Interconnecting RS232 Cable Wiring Diagram...
Страница 146: ...E 22 Appendix E ALPHABETICAL LISTING OF COMMANDS...
Страница 154: ...Index 6 INDEX...