TREATMENT  

PRG 

 

HANDLE POSITION  

 

FREQ 

DUTY CYCLE 

TIME 

HEAD 

SUGGESTED INTENSITY 

APPLICATIONS 

NUMBERS 

Acne  

U-01/10 

Affected area  

3MHz 

30% 

15 min 

5 cmq 

1,5W/cm² 

Free  

Muscle fatigue  

U-01/10 

Affected area 

1MHz 

70% 

20 min. 

5 cmq 

2 W/cm² 

2-3 

Algodystrophy 

U-01/10 

Affected area 

1MHz 

50% 

10 min. 

5 cmq 

1.0W/cm² - 1,5W/cm² 

10-15 

Anti-inflammatory 

U-01/10 

Affected area 

1MHz 

50% 

15 min. 

5 cmq 

1.0W/cm² - 1,5W/cm² 

10-15 

Arthritis 

U-01/10 

Affected area 

1MHz 

50% 

15 min. 

5 cmq 

1.0W/cm² - 1,5W/cm² 

10-15 

Fingers arthritis 

U-01/10 

Hand fingers  

1MHz 

40% 

15 min. 

1 cmq 

1.5W/cm² - 2W/cm² 

10-15 

Arthrosis 

U-01/10 

Affected area 

1MHz 

50% 

15 min. 

5 cmq 

1.5W/cm² - 2W/cm² 

10-15 

Bursitis 

U-01/10 

Affected area 

1MHz 

30% 

15 min. 

5 cmq 

2W/cm² 

10-15 

Brachialgia 

U-01/10 

Trapezium and arm 

1MHz 

30% 

15 min. 

5 cmq 

2W/cm² 

10-15 

Capsulitis  

U-01/10 

Shoulder  

1MHz 

30% 

15 min. 

5 cmq 

2W/cm² 

10-15 

Cavitations 

U-01/10 

Affected area 

1MHz 

70% 

20 min. 

5 cmq 

2W/cm² - 3W/cm² 

20-30 

T-T headache 

U-01/10 

Cervical area  

1MHz 

50% 

15 min. 

5 cmq 

1.0W/cm² - 1,5W/cm² 

10-15 

T-T headache 

U-01/10 

Massetere  

1MHz 

50% 

15 min. 

5 cmq 

1.5W/cm² 

10-15 

Cervicalgias 

U-01/10 

Cervical area  

1MHz 

50% 

15 min. 

5 cmq 

1.0W/cm² - 1,5W/cm² 

10-15 

Whiplash 

U-01/10 

Cervical and  

front zone  

1MHz 

50% 

15 min. 

5 cmq 

1.0W/cm² - 1,5W/cm² 

10-15 

Condropathy 

U-01/10 

Affected area 

1MHz 

60% 

15 min. 

5 cmq 

1.0W/cm² - 1,5W/cm² 

10-15 

Muscle contractures  

U-01/10 

Affected area 

1MHz 

70% 

20 min. 

5 cmq 

2W/cm² 

4-6 

Содержание UT1

Страница 1: ...MNPG116 Rev 0 01 08 2013...

Страница 2: ...OF USE and EQUIPMENT 5 DEVICE DESCRIPTION 6 INSTALLATION 9 PRELIMINARY OPERATIONS 9 ULTRASOUND TREATMENT 10 CLEANING MAINTENANCE AND STORAGE 16 OPERATION TROUBLES 16 ENVIRONMENT PROTECTION 17 TECHNIC...

Страница 3: ...a later date on the basis of acquired experience The manufacturer reserves the right to update the production and the manuals with no obligation to update previous versions The manufacturer declines...

Страница 4: ...r radiations should not be treated with ultrasounds Using ultrasounds on the stellate ganglion on the spinal column after a laminectomy on the area surrounding the main nerves or the cranium should be...

Страница 5: ...cated for several chronic and sub chronic treatments as Muscle pains and contractures Contractures Capsulitis Bursitis Myositis Soft tissues diseases Tendinitis Tendinosis Use hospital and domestic us...

Страница 6: ...IO ON N 1 Program selection button 2 Frequency selection button 1 3MHz 3 Duty cycle selection button 4 Timer selection button 5 LCD display 6 STOP button 7 Intensity selection knob and PAUSE 8 Handle...

Страница 7: ...am indicator 2 Frequency indicator 1 3MHz 3 Duty cycle indicator 4 Timer indicator 5 Output intensity power indicator 6 Ultrasound head detector 7 PAUSE therapy indicator 1 Ultrasound head 2 LED for u...

Страница 8: ...und handle WEE Regulations Applied part type BF Product in compliance with Directive 93 42 EEC MDD Ultrasound intensity Ultrasound applicator state contact head skin Indicator of connection socket of...

Страница 9: ...is found an error code will appear on the display please read the paragraph Operation troubles to get more details Before starting treatment please pay attention to the following suggestions Put the...

Страница 10: ...E conductive gel CE 6 Regulate the intensity of the treatment using the knob 7 Press any of the PROGRAM FREQUEN DUTY CYCLE or TIME buttons during the treatment to visualise W Watts or W cm Watt sq cm...

Страница 11: ...re the only parts protected against water damage with a IPX7 grade ATTENTION Never apply the gel to the ultrasound head The treatment head will register this as contact and may emit ultrasound energy...

Страница 12: ...cted area 1MHz 50 15 min 5 cmq 1 5W cm 2W cm 10 15 Bursitis U 01 10 Affected area 1MHz 30 15 min 5 cmq 2W cm 10 15 Brachialgia U 01 10 Trapezium and arm 1MHz 30 15 min 5 cmq 2W cm 10 15 Capsulitis U 0...

Страница 13: ...le pain U 01 10 Affected area 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Rheumatic pain U 01 10 Affected area 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Dorsalgy U 01 10 Dorsal area 1MHz 50 15 min 5 c...

Страница 14: ...balgy U 01 10 Internal thigh upper zone 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Radiculitis U 01 10 Affected area 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Muscle recovery U 01 10 Affected area 1M...

Страница 15: ...ion U 01 10 Affected area 1MHz 60 15 min 5 cmq 2W cm Free Indications regarding intensity and number of sessions can vary depending on the opinion of your personal doctor or therapist In particular in...

Страница 16: ...more information about the original accessories and spare parts contact I A C E R Srl authorized centers After cleaning the external box dry all of the parts carefully before turning on the device Do...

Страница 17: ...nsions 143mmx73mmx40mm Device Dimensions 250mmx185mmx82mm Environmental conditions for use Temperature from 10 C to 40 C Relative humidity 30 85 Atmospheric pressure 800 1060hPa Environmental conditio...

Страница 18: ...should not be placed next to or on top of other devices Should it prove necessary to place it next to or on top of other devices supervision is essential at all times to control its normal functionin...

Страница 19: ...Electromagnetic environment guidance Conducted RF lEC 61000 4 6 3V effective from 150kHz to 80MHz 3V V1 Portable and mobile RF communications equipment should be used no closer to any part of the I TE...

Страница 20: ...y starting from the purchase date covering electric and electronic parts for household use In case of purchase from professional operators purchase with an invoice the warranty is 12 months All of the...

Страница 21: ...I A C E R Srl 21 I TECH UT1 All rights riserve I TECH UT1 and logo are owned by IACER and are registered...

Страница 22: ...I A C E R Srl 22...

Страница 23: ...I A C E R Srl 23...

Страница 24: ...iva e magazzino 30030 MARTELLAGO VE Via S Pertini 24 A Tel 0039 041 5401356 Fax 0039 041 5402684 Cod Fisc P IVA IT 00185480274 R E A VE N 120250 M VE001767 e mail iacer iacer it Internet www itechmedi...

Отзывы: