*1:  Field  strengths  from  fixed  transmitters,  such  as  base  stations  for  radio  (  cellular  /  cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast 

cannot  be  predicted  theoretically  with  accuracy.  To  assess  the  electromagnetic  environment  du  to 
fixed  RF  transmitters,  an  electromagnetic  site  survey  should  be  considered.  If  the  measured  field 
strength in the location in which the I-TECH UT1 device is used exceeds the applicable RF compliance 
level  above,  should  be  observed  to  verify  normal  operation.  If  abnormal  performance  is  observed, 
additional measures may be necessary, such as reorienting or relocating the I-TECH UT1. 

*2: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V ] V/m. 

 

 

Recommended separation distances between portable and mobile 

rf communications equipment and the em devices that are not intended for life support  

 

The  I-TECH  UE  device  is  intended  for  use  in  an  electromagnetic  environment  in  which  radiated  RF 
disturbances  are  controlled.  The  customer  or  the  user  of  the  I-TECH  UE  device  can  help  prevent 
electromagnetic  interference  by  maintaining  a  minimum  distance  between  portable  and  mobile  RF 
communications  equipment  (transmitters)  and  the  I-TECH  UE    as  recommended  below,  according  to 
the maximum output power of the communications equipment.  

Rated  maximum  output 
power of transmitter (W) 

Separation distance according to frequency of transmitter (m) 

From 150kHz to 

80MHz d = 1,2 

⋅⋅⋅⋅√√√√

From 80MHz to 

800MHz d = 1,2 

⋅⋅⋅⋅√√√√

From 800MHz to 2GHz 

d = 2,3 

⋅⋅⋅⋅√√√√

P

 

0.01 

0.12 

0.12 

0.23 

0.1 

0.37 

0.37 

0.74 

1.17 

1.17 

2.33 

10 

3.69 

3.69 

7.38 

100 

11.67 

11.67 

23.33 

For  transmitters  rated  at  a  maximum  output  power  not  listed  above,  the  recommended  separation 
distance 

in  meters  (m)  can  be  estimated  using  the  equation  applicable  to  the  frequency  of  the 

transmitter, where P is the maximum output power rating of the transmitter in watts (W) accordable 
to the transmitter manufacturer. 
NOTE  1:  
At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies. 
NOTE  2:  
T h e s e g u i d e l i n e s m a y n o t a p p l y i n a l l s i t u a t i o n s . E l e c t r o m a g n e t i c p r o p a g 
a t i o n i s a f f e c t e d b y a b s o r p t i o n a n d r e f l e c t i o n f r o m s t r u c t u r e s , o b j e c t s a n 
d p e o p l e. 

 

 

W

W

A

A

R

R

R

R

A

A

N

N

T

T

Y

Y

 

 

The  device  has  a  2  year  warranty  starting  from  the  purchase  date  covering 
electric  and  electronic  parts  for  household  use.  In  case  of  purchase  from 
professional operators (purchase with an invoice), the warranty is 12 months. 
All  of  the  parts  subject  to  normal  wear  and  tear  (ultrasound  head)  are  not 
covered  by  warranty  as  well  as  all  the  parts  that  are  defective  due  to  
negligence,  improper  maintenance,  tampering  or  repair  work  carried  out  by 
personnel that has not been authorised by the manufacturer or the authorised 
dealer. 

 

The warranty conditions are those described under “Warranty Regulations”. 

In accordance with the Medical Devices Directive 93/42/EEC, the manufacturer 
must be able to trace the devices at any time in order to intervene promptly in 
case of manufacturing faults.  

In the event of future repairs under warranty, the equipment must be packaged 
to prevent damage during transport and sent to the manufacturer together with 
all  of  the  accessories.  The  purchaser  only  has  the  right  to  repair  under 
warranty  when  the  equipment  is  returned  to  the  manufacturer  complete  with 
the  receipt  or  invoice  proving  the  correct  origin  of  the  product  and  purchase 
date.  

Warranty regulations. 

1.  In  the  event  f  repairs  under  warranty,  the  purchaser  must  include  in  the  package  the  receipt  or 

invoicing proving the purchase date.  

2.  The  electronic  parts  are covered  by  a  24-months  (12 months  for  professional  user)  warranty.  The 

warranty is given through the point of sale or directly from the manufacturer

3.  The warranty covers exclusively product damage causing operational defects. 
4.  
The  warranty  covers  exclusively  the  repair  or  replacement  free  of  charge,  including  labour,  of 

components found to be defective in terms of manufacture or material. 

5.  The  warranty  does  not  apply  to  damage  caused  by  neglect  or  use  not  complying  with  the 

instructions  provided,  damage  caused  by  work  carried  out  by  unauthorized  personnel  or  damage 
caused by accidental causes or the buyer’s negligence, with particular reference to external parts. 

6.  The  warranty  does  also  not  apply  to  damage  to  the  equipment  caused  by  incompatible  power 

supplies 

7.

 

Parts subject to wear after use are excluded from the warranty.

 

8.  The warranty does not include transport costs to be paid by the purchaser in relation to the method 

and speed of transport. 

9.  The warranty empire after 24 months (12 months for professional user).. After such time repair work 

will be carried out at the rates currently in force for the parts replaced and the labour and transport 
costs.  

10.  Any controversy will fall within the exclusive jurisdiction of the Venice courts
  

Содержание UT1

Страница 1: ...MNPG116 Rev 0 01 08 2013...

Страница 2: ...OF USE and EQUIPMENT 5 DEVICE DESCRIPTION 6 INSTALLATION 9 PRELIMINARY OPERATIONS 9 ULTRASOUND TREATMENT 10 CLEANING MAINTENANCE AND STORAGE 16 OPERATION TROUBLES 16 ENVIRONMENT PROTECTION 17 TECHNIC...

Страница 3: ...a later date on the basis of acquired experience The manufacturer reserves the right to update the production and the manuals with no obligation to update previous versions The manufacturer declines...

Страница 4: ...r radiations should not be treated with ultrasounds Using ultrasounds on the stellate ganglion on the spinal column after a laminectomy on the area surrounding the main nerves or the cranium should be...

Страница 5: ...cated for several chronic and sub chronic treatments as Muscle pains and contractures Contractures Capsulitis Bursitis Myositis Soft tissues diseases Tendinitis Tendinosis Use hospital and domestic us...

Страница 6: ...IO ON N 1 Program selection button 2 Frequency selection button 1 3MHz 3 Duty cycle selection button 4 Timer selection button 5 LCD display 6 STOP button 7 Intensity selection knob and PAUSE 8 Handle...

Страница 7: ...am indicator 2 Frequency indicator 1 3MHz 3 Duty cycle indicator 4 Timer indicator 5 Output intensity power indicator 6 Ultrasound head detector 7 PAUSE therapy indicator 1 Ultrasound head 2 LED for u...

Страница 8: ...und handle WEE Regulations Applied part type BF Product in compliance with Directive 93 42 EEC MDD Ultrasound intensity Ultrasound applicator state contact head skin Indicator of connection socket of...

Страница 9: ...is found an error code will appear on the display please read the paragraph Operation troubles to get more details Before starting treatment please pay attention to the following suggestions Put the...

Страница 10: ...E conductive gel CE 6 Regulate the intensity of the treatment using the knob 7 Press any of the PROGRAM FREQUEN DUTY CYCLE or TIME buttons during the treatment to visualise W Watts or W cm Watt sq cm...

Страница 11: ...re the only parts protected against water damage with a IPX7 grade ATTENTION Never apply the gel to the ultrasound head The treatment head will register this as contact and may emit ultrasound energy...

Страница 12: ...cted area 1MHz 50 15 min 5 cmq 1 5W cm 2W cm 10 15 Bursitis U 01 10 Affected area 1MHz 30 15 min 5 cmq 2W cm 10 15 Brachialgia U 01 10 Trapezium and arm 1MHz 30 15 min 5 cmq 2W cm 10 15 Capsulitis U 0...

Страница 13: ...le pain U 01 10 Affected area 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Rheumatic pain U 01 10 Affected area 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Dorsalgy U 01 10 Dorsal area 1MHz 50 15 min 5 c...

Страница 14: ...balgy U 01 10 Internal thigh upper zone 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Radiculitis U 01 10 Affected area 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Muscle recovery U 01 10 Affected area 1M...

Страница 15: ...ion U 01 10 Affected area 1MHz 60 15 min 5 cmq 2W cm Free Indications regarding intensity and number of sessions can vary depending on the opinion of your personal doctor or therapist In particular in...

Страница 16: ...more information about the original accessories and spare parts contact I A C E R Srl authorized centers After cleaning the external box dry all of the parts carefully before turning on the device Do...

Страница 17: ...nsions 143mmx73mmx40mm Device Dimensions 250mmx185mmx82mm Environmental conditions for use Temperature from 10 C to 40 C Relative humidity 30 85 Atmospheric pressure 800 1060hPa Environmental conditio...

Страница 18: ...should not be placed next to or on top of other devices Should it prove necessary to place it next to or on top of other devices supervision is essential at all times to control its normal functionin...

Страница 19: ...Electromagnetic environment guidance Conducted RF lEC 61000 4 6 3V effective from 150kHz to 80MHz 3V V1 Portable and mobile RF communications equipment should be used no closer to any part of the I TE...

Страница 20: ...y starting from the purchase date covering electric and electronic parts for household use In case of purchase from professional operators purchase with an invoice the warranty is 12 months All of the...

Страница 21: ...I A C E R Srl 21 I TECH UT1 All rights riserve I TECH UT1 and logo are owned by IACER and are registered...

Страница 22: ...I A C E R Srl 22...

Страница 23: ...I A C E R Srl 23...

Страница 24: ...iva e magazzino 30030 MARTELLAGO VE Via S Pertini 24 A Tel 0039 041 5401356 Fax 0039 041 5402684 Cod Fisc P IVA IT 00185480274 R E A VE N 120250 M VE001767 e mail iacer iacer it Internet www itechmedi...

Отзывы: