I.A.C.E.R. Srl 

11

 

 

Put a quantity of conductive gel on the area to be treated. Move the ultrasound head 
during  therapy  session  in  a  circular  motion.  The  treated  area  should  be  twice  the 
ultrasound head area.  

 

If  the  body  surface  is  very  irregular,  making  it  difficult  to  obtain  good  contact 
between the ultrasound head and the body, or if direct contact must be avoided (e.g. 
due to pain), the affected area may be treated under water (subaqual method). The 
water should be degassed (by previous boiling ) in order to prevent air bubbles that 
could decrease the effectiveness of the treatment.  

ATTENTION.  The  handpiece and  its  cable  are  the  only  parts  protected  against  water 
damage with a IPX7 grade.  

ATTENTION.    Never  apply  the  gel  to  the  ultrasound  head.  The  treatment  head  will 
register  this  as  contact  and  may  emit  ultrasound  energy,  which  could  damage  the 
ultrasound  head.  Always  use  the  gel  certificated  with  the  requirements  of  the 
medical, such as with  CE mark.  
 

 

 

 

 

 

 

 

 

 

 

Programs features and main applications.   

Make reference to the following table for programs features. All parameters are 

adjustable by the user.   

PROG. 

FREQ. 

DUTY CYCLE 

TIME 

SUGGESTED INT. 

U-01 

1MHz 

80% 

10 min. 

1.0W/cm² 

U-02 

1MHz 

50% 

10 min. 

1.0W/cm² 

U-03 

1MHz 

50% 

20 min. 

1.5W/cm² 

U-04 

1MHz 

50% 

15 min. 

1.0W/cm² 

1.5W/cm² 

2.0W/cm² 

U-05 

3MHz 

80% 

15 min. 

1.0W/cm² 

U-06 

1MHz 

30% 

15 min. 

1.5W/cm² 

U-07 

1MHz 

80% 

15 min. 

1.0W/cm² 

1.5W/cm² 

U-08 

1MHz 

80% 

8 min. 

1.5W/cm² 

U-09 

1MHz 

50% 

12 min. 

1.5W/cm² 

U-10 

3MHz 

80% 

10 min. 

1.0W/cm² 

 

 

 

Содержание UT1

Страница 1: ...MNPG116 Rev 0 01 08 2013...

Страница 2: ...OF USE and EQUIPMENT 5 DEVICE DESCRIPTION 6 INSTALLATION 9 PRELIMINARY OPERATIONS 9 ULTRASOUND TREATMENT 10 CLEANING MAINTENANCE AND STORAGE 16 OPERATION TROUBLES 16 ENVIRONMENT PROTECTION 17 TECHNIC...

Страница 3: ...a later date on the basis of acquired experience The manufacturer reserves the right to update the production and the manuals with no obligation to update previous versions The manufacturer declines...

Страница 4: ...r radiations should not be treated with ultrasounds Using ultrasounds on the stellate ganglion on the spinal column after a laminectomy on the area surrounding the main nerves or the cranium should be...

Страница 5: ...cated for several chronic and sub chronic treatments as Muscle pains and contractures Contractures Capsulitis Bursitis Myositis Soft tissues diseases Tendinitis Tendinosis Use hospital and domestic us...

Страница 6: ...IO ON N 1 Program selection button 2 Frequency selection button 1 3MHz 3 Duty cycle selection button 4 Timer selection button 5 LCD display 6 STOP button 7 Intensity selection knob and PAUSE 8 Handle...

Страница 7: ...am indicator 2 Frequency indicator 1 3MHz 3 Duty cycle indicator 4 Timer indicator 5 Output intensity power indicator 6 Ultrasound head detector 7 PAUSE therapy indicator 1 Ultrasound head 2 LED for u...

Страница 8: ...und handle WEE Regulations Applied part type BF Product in compliance with Directive 93 42 EEC MDD Ultrasound intensity Ultrasound applicator state contact head skin Indicator of connection socket of...

Страница 9: ...is found an error code will appear on the display please read the paragraph Operation troubles to get more details Before starting treatment please pay attention to the following suggestions Put the...

Страница 10: ...E conductive gel CE 6 Regulate the intensity of the treatment using the knob 7 Press any of the PROGRAM FREQUEN DUTY CYCLE or TIME buttons during the treatment to visualise W Watts or W cm Watt sq cm...

Страница 11: ...re the only parts protected against water damage with a IPX7 grade ATTENTION Never apply the gel to the ultrasound head The treatment head will register this as contact and may emit ultrasound energy...

Страница 12: ...cted area 1MHz 50 15 min 5 cmq 1 5W cm 2W cm 10 15 Bursitis U 01 10 Affected area 1MHz 30 15 min 5 cmq 2W cm 10 15 Brachialgia U 01 10 Trapezium and arm 1MHz 30 15 min 5 cmq 2W cm 10 15 Capsulitis U 0...

Страница 13: ...le pain U 01 10 Affected area 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Rheumatic pain U 01 10 Affected area 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Dorsalgy U 01 10 Dorsal area 1MHz 50 15 min 5 c...

Страница 14: ...balgy U 01 10 Internal thigh upper zone 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Radiculitis U 01 10 Affected area 1MHz 50 15 min 5 cmq 1 0W cm 1 5W cm 10 15 Muscle recovery U 01 10 Affected area 1M...

Страница 15: ...ion U 01 10 Affected area 1MHz 60 15 min 5 cmq 2W cm Free Indications regarding intensity and number of sessions can vary depending on the opinion of your personal doctor or therapist In particular in...

Страница 16: ...more information about the original accessories and spare parts contact I A C E R Srl authorized centers After cleaning the external box dry all of the parts carefully before turning on the device Do...

Страница 17: ...nsions 143mmx73mmx40mm Device Dimensions 250mmx185mmx82mm Environmental conditions for use Temperature from 10 C to 40 C Relative humidity 30 85 Atmospheric pressure 800 1060hPa Environmental conditio...

Страница 18: ...should not be placed next to or on top of other devices Should it prove necessary to place it next to or on top of other devices supervision is essential at all times to control its normal functionin...

Страница 19: ...Electromagnetic environment guidance Conducted RF lEC 61000 4 6 3V effective from 150kHz to 80MHz 3V V1 Portable and mobile RF communications equipment should be used no closer to any part of the I TE...

Страница 20: ...y starting from the purchase date covering electric and electronic parts for household use In case of purchase from professional operators purchase with an invoice the warranty is 12 months All of the...

Страница 21: ...I A C E R Srl 21 I TECH UT1 All rights riserve I TECH UT1 and logo are owned by IACER and are registered...

Страница 22: ...I A C E R Srl 22...

Страница 23: ...I A C E R Srl 23...

Страница 24: ...iva e magazzino 30030 MARTELLAGO VE Via S Pertini 24 A Tel 0039 041 5401356 Fax 0039 041 5402684 Cod Fisc P IVA IT 00185480274 R E A VE N 120250 M VE001767 e mail iacer iacer it Internet www itechmedi...

Отзывы: