Put a quantity of conductive gel on the area to be treated. Move the ultrasound head
during therapy session in a circular motion. The treated area should be twice the
ultrasound head area.
If the body surface is very irregular, making it difficult to obtain good contact
between the ultrasound head and the body, or if direct contact must be avoided (e.g.
due to pain), the affected area may be treated under water (subaqual method). The
water should be degassed (by previous boiling ) in order to prevent air bubbles that
could decrease the effectiveness of the treatment.
ATTENTION. The handpiece and its cable are the only parts protected against water
damage with a IPX7 grade.
ATTENTION. Never apply the gel to the ultrasound head. The treatment head will
register this as contact and may emit ultrasound energy, which could damage the
ultrasound head. Always use the gel certificated with the requirements of the
medical, such as with CE mark.
Programs features and main applications.
Make reference to the following table for programs features. All parameters are
adjustable by the user.
PROG.
FREQ.
DUTY CYCLE
TIME
SUGGESTED INT.
U-01
1MHz
80%
10 min.
1.0W/cm²
U-02
1MHz
50%
10 min.
1.0W/cm²
U-03
1MHz
50%
20 min.
1.5W/cm²
U-04
1MHz
50%
15 min.
1.0W/cm²
1.5W/cm²
2.0W/cm²
U-05
3MHz
80%
15 min.
1.0W/cm²
U-06
1MHz
30%
15 min.
1.5W/cm²
U-07
1MHz
80%
15 min.
1.0W/cm²
1.5W/cm²
U-08
1MHz
80%
8 min.
1.5W/cm²
U-09
1MHz
50%
12 min.
1.5W/cm²
U-10
3MHz
80%
10 min.
1.0W/cm²