8. Troubleshooting
8-1
i-Smart 300 Blood Gas Analyzer
Troubleshooting
Guideline
If the problems described below are encountered during
installation and/or operation of the analyzer, try the suggested
solutions as described in this manual.
If the problem persists, please call a service engineer for further
assistance.
Barcode
Scanner
When scanning barcode, the scanner does not emit the red light:
1.
Move the barcodes to be scanned in front of the marking
“ BARCODE SCAN”.
2.
If there does not show a red light beam, press the “SCAN”
button on the rear of the analyzer by a sharp object like a needle.
Then, retry to read the barcodes. It may not be able to read the
barcodes at a dark place. In this case, please enhance the
brightness of the light in around.
3.
If the same problem occurs, turn off and on the power switch of
the analyzer.
Screen
If one of following situations applies
:
The screen does not respond.
The screen is frozen.
Abnormal screen appears.
1.
Turn off the power switch of the analyzer.
2.
Wait 10 seconds.
3.
Turn on the power switch of the analyzer.
Содержание i-Smart 300
Страница 2: ...Operators should read the entire manual before installing and operating the analyzer...
Страница 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Страница 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Страница 32: ...3 Analyzer Setting 3 7 i Smart 300 Screen Menus continued Keyboards continued Numerical Keyboard...
Страница 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Страница 63: ...5 Database 5 Database Blood Sample Data 5 1 QC Data 5 5 Calibration Data 5 9 Cartridge Data 5 12...
Страница 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Страница 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Страница 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...