Appendices
Operator’s Manual
10-1
Appendix A: Sample Collection and Handling
Caution
Observe the general cautions required of the hospital, laboratory,
or other testing facility.
All biohazardous materials should be handled and disposed of in
accordance with applicable rules and regulations.
The blood and/or collecting tools, and used cartridges should be
treated as biohazardous materials.
Wear appropriate personal protective clothing (lab coat, gloves,
goggles, etc.). And Be careful not to let the blood and/or
collecting tools directly touch the mouth, eyes, mucus
membranes, or any area with broken skin.
Wash hands after collecting the sample or using the analyzer.
Sample
preparation
Use whole blood to a blood gas test.
To draw blood samples, use containers (syringe, capillary, blood
collection tube) treated with heparin. We recommend using a
dry-sprayed heparin tube or a solid heparin tube. We do not
recommend using a tube treated with anticoagulants such as
Citrate, EDTA, Oxalate, NaF, etc., except heparin, since the
blood-sampling tube containing such anticoagulants besides
heparin can drop the performance of the sensors and/or can
influence the analysis result.
Make sure there are no small air bubbles trapped in the collected
sample. Small bubbles in the sample can affect the results.
Samples containing clots and/or samples in which hemolysis is
present or suspected must not be used.
Use whole blood samples for blood gas analysis should be
analyzed as soon as possible. If immediately analysis is not
possible, the sealed blood container must be placed in a bath
containing ice and water. Ice-water stored samples at 1 to 4°C
may give reliable results for up to 30 minutes.
Содержание i-Smart 300
Страница 2: ...Operators should read the entire manual before installing and operating the analyzer...
Страница 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Страница 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Страница 32: ...3 Analyzer Setting 3 7 i Smart 300 Screen Menus continued Keyboards continued Numerical Keyboard...
Страница 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Страница 63: ...5 Database 5 Database Blood Sample Data 5 1 QC Data 5 5 Calibration Data 5 9 Cartridge Data 5 12...
Страница 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Страница 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Страница 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...