7. Maintenance
Operator’s Manual
7-2
Calibration,
continued
Cal 1 Auto
Repeat
If the drift error of a sensor has occurred in the previous Cal 1,
the analyzer will automatically repeat Cal 1 up to three times.
While the Cal 2 repeat is in progress, the “
Drift error has
occured. Cal 1 is in progress. Please wait...
” message will
appear.
Cal 2 Auto
Repeat
If the slope error of a sensor has occurred in the previous Cal 2,
the analyzer will automatically repeat Cal 2 up to two times.
While the Cal 2 repeat is in progress, the “
Slope error has
occured. Cal 2 is in progress. Please wait...
” message will
appear.
If the same sensor fails in Cal 2 three times in succession, the
sensor state will appear as on the main screen.
Содержание i-Smart 300
Страница 2: ...Operators should read the entire manual before installing and operating the analyzer...
Страница 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Страница 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Страница 32: ...3 Analyzer Setting 3 7 i Smart 300 Screen Menus continued Keyboards continued Numerical Keyboard...
Страница 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Страница 63: ...5 Database 5 Database Blood Sample Data 5 1 QC Data 5 5 Calibration Data 5 9 Cartridge Data 5 12...
Страница 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Страница 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Страница 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...