6. Shutdown
6-3
i-Smart 300 Blood Gas Analyzer
Cartridge Restart
Power Recovery
When the analyzer recovers from power interruption, the
cartridge inside the analyzer can be continuously used, only if:
The analyzer was in Ready state or in process of calibration
when the power was interrupted and the power returns
within 1 hour.
The analyzer was analyzing a sample when the power was
interrupted and the power returns within 20 minutes.
The analyzer was analyzing a QC sample when the power
was interrupted and the power returns within 1 hour.
1.
When the analyzer recovers from power interruption with the
cartridge inside, one of following
Cartridge Restart
screens
will appear.
2.
The
Cartridge Restart
from power recovery will take about 10
minutes.
3.
After the
Cartridge Restart
, the analyzer will return to the
Ready state.
Содержание i-Smart 300
Страница 2: ...Operators should read the entire manual before installing and operating the analyzer...
Страница 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Страница 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Страница 32: ...3 Analyzer Setting 3 7 i Smart 300 Screen Menus continued Keyboards continued Numerical Keyboard...
Страница 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Страница 63: ...5 Database 5 Database Blood Sample Data 5 1 QC Data 5 5 Calibration Data 5 9 Cartridge Data 5 12...
Страница 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Страница 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Страница 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...