3. Analyzer Setting
사용 설명서
3-12
Sample Setup,
continued
Critical Limits
1.
Go to
Menu > Setup
(
Setup
menu is protected by password.)
>
Critical Limits Setup
. The following screen will appear.
2.
Press each box and enter low and high values of the critical
limits range for each parameter.
Note:
To delete the entered data,
press
Current Tab Clear
.
Before saving the setup, the analyzer will validate the entered
values to confirm that:
-
Values for both low and high limits are entered.
-
Low limit value is lower than high limit value.
If any of the entered values are not valid, the analyzer will not
save the setup and ask to correct the invalid value at the setup
screen.
3.
Press
Save
to save the setup and go back to the main screen.
To exit to main screen without saving the changed setting, press
Close
.
Содержание i-Smart 300
Страница 2: ...Operators should read the entire manual before installing and operating the analyzer...
Страница 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Страница 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Страница 32: ...3 Analyzer Setting 3 7 i Smart 300 Screen Menus continued Keyboards continued Numerical Keyboard...
Страница 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Страница 63: ...5 Database 5 Database Blood Sample Data 5 1 QC Data 5 5 Calibration Data 5 9 Cartridge Data 5 12...
Страница 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Страница 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Страница 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Страница 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...