Within Run Precision
Blood average 39.5 mg/dL (2.2 mmol/L)
SD=1.2 mg/dL
(0.07 mmol/L)
Blood average 76.7 mg/dL (4.3 mmol/L)
SD=1.6 mg/dL
(0.1 mmol/L)
Blood average 128.9 mg/dL (7.2 mmol/L) CV=1.7%
Blood average 203.5 mg/dL (11.3 mmol/L) CV=1.7%
Blood average 298.1 mg/dL (16.6 mmol/L) CV=1.9%
Total Precision
Control average 44.4 mg/dL (2.5 mmol/L)
1.9 mg/dL
(0.11 mmol/L)
Control average 136.1 mg/dL (7.6 mmol/L) 3.5%
Control average 356.4 mg/dL (19.8 mmol/L) 4.5%
Error Display Messages and Troubleshooting
Test strip has been
used.
Use a new strip.
Blood sample applied
before
symbol
appeared.
Repeat test with a
new strip.
Wait until
symbol
appears before
applying sample.
Test has been
performed outside
the operating
temperature range.
Take system where
temperature is
within operating
range (10-40 /50-
104 ) and repeat
test after 30
minutes.
Message
Possible Cause
Solution
Performance Evaluation
Precision: Precision studies were performed in a laboratory using
the e-CheKer BGM System.
This study shows that there could be variation of up to 4.5%