Setting the Meter

Setting the Date

Press 

C

 button until blinking number 

represents correct date. 
Press M button to set and move to time setting 

mode. 

Helpful Tip
Press and hold the 

M

 button for three (3) seconds to begin 

and end the setting of the date, time, year or measurement 

unit on the meter. Use the 

C

 button for scrolling. To scroll 

more quickly, hold the 

C

 button down.

Step 3

Setting the Time

The meter can be set in an AM/PM 12-hour or 

24-hour format. Press 

C

 button to select 

format.

Step 4

Setting the Hour

Press 

M

 button to move to hour setting mode. 

Press 

C

 button until correct hour appears. After 

hour is set, press 

M

 button to enter minute 

setting mode.

Step 5

Setting the Minute

Press 

C

 button until correct minute 

appears. After minute is set, press 

M

 

button to move to year setting mode.

Step 2

Setting the Year

The year will appear on the display when 

entering  year  setting  mode  for  the  first 

time.  Press 

C

  button  until  correct  year 

appears.  After  setting  the  year,  press 

M

 

button to move to unit setting mode.

Setting the Measurement Unit

Press 

C

 button to change the unit to 

mmol/L or mg/dL. After selecting 

measurement unit, press 

M

 button for 

three (3) seconds to exit setting mode.

Step 6

Step 7

Helpful Tip
The mg/dL is a standard unit in the United States. 

The most commonly used unit in Australia, Canada, China 

and in some European countries is mmol/L.

Setting the Time

Setting the Year & Measurement Unit

Содержание e-Checker

Страница 1: ...6 Wolgye4 dong Nowon gu Seoul 139 845 Korea www i sens com Medical Technology Promedt Consulting GmbH Altenhofstrasse 80 D 66386 St Ingbert Germany 2012 i SENS Inc All Rights Reserved PGB6E3001 REV4x...

Страница 2: ...ur purchase of the e CheKer Blood Glucose Monitoring System The system provides you with convenient safe and less painful blood glucose in vitro i e outside the body diagnostic monitoring You ll recei...

Страница 3: ...e 9 Setting the Time 10 Setting the Year and Measurement Unit 11 Setting the Sound On OFF 12 StoringTest Results in Meter Memory 13 Deleting StoredTest Results from Meter Memory 14 e CheKer Blood Gluc...

Страница 4: ...g along with alternate sites testing on forearm palm thigh and calf The e CheKer blood glucose monitoring system is intended for self testing outside the body invitro diagnostic use The glucose in the...

Страница 5: ...y or setting modes Meter Display Test result symbol and simple message display C Button Power button to review previous test results or change code date time year or measurement units Test Strip Port...

Страница 6: ...L DecimalPoint Appears when measurement unit issettommol L unit Blood Drop Symbol Indicates meter is ready to receive blood sample Month Day Hour Minute code Shows code for three 3 seconds after test...

Страница 7: ...ter hour is set press M button to enter minute setting mode Step 5 Setting the Minute Press C button until correct minute appears After minute is set press M button to move to year setting mode Step 2...

Страница 8: ...ress C button to turn meter off Helpful Tip Each time you press the C button you ll see previous test results dates and times displayed in chronological order To see test results displayed more quickl...

Страница 9: ...es or No To delete all memory choose Yes Press M button All past stored test results and their dates and times will be deleted The number zero will appear Or if you do not wish to delete the memory ch...

Страница 10: ...se test strips Keep test strips and the test strip bottle away from children The test strips and bottle cap may be choking hazards Drying agents in the bottle cap may be harmful if inhaled or swallowe...

Страница 11: ...test strips The code number on the display and test strip bottle must match If the meter is not correctly coded you will receive inaccurate results For accurate results the code numbers on the e CheKe...

Страница 12: ...er comes first Only use e CheKer Control Solution to test e CheKer Test Strips and the e CheKer Meter e CheKer Control Solution It contains a known amount of glucose that reacts with a e CheKer Test S...

Страница 13: ...t Gently push test strip into the port until meter beeps A code number will appear for three 3 seconds The symbol will show on the display Confirm matching code numbers on display and test strip bottl...

Страница 14: ...same sample site may cause pain or skin calluses Choose a different site each time you test Out of Range Result Factors If your control solution test results do not fall within the range printed on th...

Страница 15: ...le at different sample sites The lancing device s adjustable tip allows the best depth of skin penetration for an adequate sample size A beginning setting of three 3 is recommended Step 4 Step 5 Prepa...

Страница 16: ...Match Test Strips on pages 17 18 Step 1 Step 2 Place cocked lancing device against fingertip side Helpful Tip Choose a different fingertip each time you retrieve a sample to minimize discomfort Abnorm...

Страница 17: ...l count down from five to one 5 1 on the display Test results time and date will appear and automatically store in meter memory Remove used test strip from port Meter will turn off after three 3 secon...

Страница 18: ...the test results from the fingertip test and AST may differ This is because things such as lifestyle and ingested food have an effect on glucose levels Things to know when using AST Alternative SiteT...

Страница 19: ...e glucose levels to reach the same value Use a fingertip for drawing if you suffer from hypoglycemia or have experienced hypoglycemic shock or symptoms If the sample drop of blood runs or spreads due...

Страница 20: ...etween 20 to 600 mg dL 1 1 33 3 mmol L Please refer to the instructions below if you receive a Hi or Lo message If you receive a Hi or Lo message repeat the test If the result is the same contact your...

Страница 21: ...r acetone An abnormally high or low red blood cell count hematocrit levels over 60 or below 20 may produce inaccurate results A normal red blood cell count is in the 30 50 range It may be below or ove...

Страница 22: ...sts Accuracy The accuracy of the e CheKer BGM System Model GM505FA was assessed by comparing blood glucose results obtained by patients with those obtained using aYSI Model 2300 Glucose Analyzer a lab...

Страница 23: ...rol average 356 4 mg dL 19 8 mmol L 4 5 Error Display Messages andTroubleshooting Teststriphasbeen used Use a new strip Bloodsampleapplied before symbol appeared Repeattestwitha newstrip Wait until sy...

Страница 24: ...e meter is on by pressing M button If meter doesn t turn on battery may be low Change the battery Be sure the battery has facing up If not reinstall correctly Meter does not countdown from five to one...

Страница 25: ...pering use or service of the meter Any claim must be made within the warranty period The i SENS company will at its discretion repair or replace a defective meter or meter part that is covered by this...

Страница 26: ...48 MEMO MEMO...

Страница 27: ...50 MEMO...

Отзывы: