28   www.i-sens.com

www.i-sens.com 29

Step 8
Obtain a blood sample using the 

lancing device. Place the device against 

the pad of the finger. The best puncture 

sites are on the middle or ring fingers. 

Press the release button. Remove the 

device from the finger. Wait a few 

seconds for a blood drop to form. You need a minimum volume 

of 0.4 microliter for blood glucose test with CareSens PRO test 

strip.  (Actual size of 0.4 μL:    )

Applying Blood Sample

Step 7
Insert a test strip with 

the contact bars facing 

upwards into the meter’s 

test strip port. Push the 

strip in gently until the 

meter beeps. Be careful 

not to bend the test strip. 

The  

  symbol will 

appear on the screen.

Preparing the Meter and Test Strip

Blood Glucose Testing

Step 5
The lancing device has six puncture 

depth settings (0 for a shallow 

puncture, 5 for a deeper puncture). 

Choose a depth by rotating the top 

portion of the adjustable tip until the 

desired number aligns with the arrow.

Step 6
To cock the lancing device, hold the 

body of lancing device in one hand and 

pull the sliding barrel with the other 

hand. The device is loaded when you 

feel a click and the load confirmation  

window turns red.

Note:

 The skin depth to get blood samples will vary for 

various people at different sample sites. The lancing device's 

adjustable tip allows the best depth of skin penetration to 

get an adequate sample size. 

Note:

 

0 = a shallow puncture for softer skin

5 = a deeper puncture for thick or calloused skin

Содержание CareSens PRO

Страница 1: ...l Rights Reserved PGL5E1006 REVA 11 2018 i SENS Inc 43 Banpo daero 28 gil Seocho gu Seoul 06646 Korea www i sens com Medical Technology Promedt Consulting GmbH Altenhofstrasse 80 66386 St Ingbert Germ...

Страница 2: ...in just 5 seconds with a small 0 4 L blood sample WelcometotheCareSens PROBloodGlucose MonitoringSystem No part of this document may be reproduced in any form or by any means without the prior writte...

Страница 3: ...cator 19 Setting the Strip Expiration Date Indicator 20 Checking the System 21 Control Solution Testing 22 Comparing the Control Solution Test Results 25 Testing Using the Lancing Device 26 Blood Gluc...

Страница 4: ...ause inaccurate results If you need assistance please contact your authorised i SENS sales representative or visit www i sens com for more information Important Information ReadThis First For optimum...

Страница 5: ...horised i SENS sales representative Specifications CareSens PRO Blood Glucose Monitoring System Operating ranges Storage Conditions Product specification Measurement range 10 600 mg dL 0 56 33 3 mmol...

Страница 6: ...meter check the battery compartment and insert batteries if empty When the symbol appears on the display while the meter is in use the batteries should be replaced as soon as possible The test results...

Страница 7: ...e CareSens PRO test strip package insert Caution Keep the meter and testing supplies away from young children Drying agents in the vial cap may be harmful if inhaled or swallowed and may cause skin or...

Страница 8: ...e after eating Fasting test flag used for tests done after fasting for at least 8 hours Month Day Hour Minute 10 12 7 8 9 11 13 14 2 3 5 4 6 10 9 8 1 14 11 12 13 Note It is recommended to check if the...

Страница 9: ...ss the or button to adjust until the correct year appears When the present year appears press the S button to confirm your selection and to go to the next step Step 3 Setting the Month A number indica...

Страница 10: ...When you press and hold the button to set the post meal PP2 alarm When it is time for a pre set blood glucose test Setting the Sound On OFF Note The symbol is displayed only when the sound is set to O...

Страница 11: ...level is before setting your level Note If the test result is lower than the pre set hypoglycemia level the meter will display the following HYPowarning willblinkatthe bottom Step 9 This setting allow...

Страница 12: ...L and or H The CareSens PRO Control Solution contains a known amount of glucose and is used to check that the meter and the test strips are working properly The CareSens PRO test strip vials have Car...

Страница 13: ...up Step 2 Activating Control Solution Test Mode Press and hold the button for 3 seconds to activate the Control SolutionTest Mode This will also flag the control solution test result To undo the cont...

Страница 14: ...trol solution test if the test result falls outside of the range Out of range results may occur in following situations If results continue to fall outside the range the test strip and meter may not b...

Страница 15: ...device The lancing device is for use by a single user only and should not be shared with anyone Use a soft cloth or tissue to wipe the lancing device If necessary a small amount of alcohol on a soft c...

Страница 16: ...The symbol will appear on the screen Preparing the Meter and Test Strip Blood GlucoseTesting Step 5 The lancing device has six puncture depth settings 0 for a shallow puncture 5 for a deeper puncture...

Страница 17: ...the blood is going in Test result will appear after the meter counts down from 5 to 1 The result will be automatically stored in the meter s memory If the test strip is removed after the test result...

Страница 18: ...st press or button to select a pre meal flag a post meal flag or a fasting flag When you remove the test strip while the desired flag is blinking the test result is stored with the flag If you do not...

Страница 19: ...However since there are many nerve endings in the fingertip it can be quite painful When doing a glucose test using different parts of the body such as the forearms and palms can reduce the pain durin...

Страница 20: ...hyperglycemia or hypoglycemia When the results of the test do not reflect your opinion retest using the fingertip test If the fingertip result still does not reflect the way you feel please consult y...

Страница 21: ...d for the last test period button button Average 1 7 14 30 60 90days Target Blood Glucose Ranges Reminders Time of day Before breakfast Before lunch or dinner 1 hour after meals 2 hours after meals Be...

Страница 22: ...ormed pre meals with the symbol for the last test period Step 4 Viewing Post meal Averages Press the button to view 1 7 14 30 and 90 day average values and the number of tests performed post meals wit...

Страница 23: ...ting the Post meal Alarm PP2 alarm Step 1 Press any button to turn the meter on The current date and time will be displayed on the bottom of the screen followed by the 1 day average value and the numb...

Страница 24: ...on again to cancel alarm1 OFF will blink on the screen Step 3 Press the button while On blinks to adjust the time of alarm1 A number indicating the hour will blink on the screen Press the button to se...

Страница 25: ...Repeat the test with a new test strip The temperature during the test was above or below the operating range Move to an area where the temperature is within the operating range 5 45 C and repeat the...

Страница 26: ...ter inserting a test strip The test does not start even after applying the blood sample on the strip The test result does not match the way you feel General Troubleshooting Note If the problem is not...

Страница 27: ...8 3 3 1 7 mg dL 0 2 0 1 mmol L 96 to 144 5 3 to 8 0 3 6 1 7 280 to 420 15 5 to 23 3 8 9 6 2 Range mg dL mmol L Average of difference Hct 15 65 Performance Characteristics The performance of CareSens...

Страница 28: ...in 15 of the medical laboratory values at glucose concentrations at or above 100 mg dL 5 55 mmol L Compounds of xylose 9 69 mg dL 0 54 mmol L at glucose concentrations of 50 100 mg dL 2 8 5 6 mmol L m...

Страница 29: ...handling tampering use or service of the meter Any claim must be made within the warranty period i SENS will at its discretion repair or replace a defective meter or meter part that is covered by this...

Отзывы: