Step 4
Waiting for the Result
The display segments will rotate
clockwise and a test result will
appear after the meter counts
down from 5 to 1.
The test result with control
solution flag is stored in the
memory but not included in the
averages.
2.0-3.8 mmol/L
Control Solution Range
Control L:
101–151 mg/dL 5.6–8.4 mmol/L
Control H:
184–276 mg/dL10.3–15.3 mmol/L
Step 5
Comparing the Result
Compare the result displayed on the
meter to the range printed on the
test strip vial. The result should fall
within the range.
Caution:
The range printed on the test strip vial is for the
Control Solution only. It has nothing to do with your blood
glucose level.
Comparing the Control Solution Test Results
The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of the range.
Out of range results may occur in following situations:
If results continue to fall outside the range, the test strip and
meter may not be working properly. Do not use your system
and contact i-SENS sales representative.
Situations
• When the control solution bottle was
not shaken well,
• When the meter, test strip, or the
control solution were exposed to high
or low temperatures,
• When the first drop of the control
solution was not discarded or the tip of
the bottle was not wiped clean,
• When the meter is not functioning
properly.
• When the control solution is past the
expiration date printed on the bottle,
• When the control solution is past its
discard date,
• When the control solution is
contaminated.
Do This
Repeat the
control solution
test by referring
to the notes on
page 21.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.
Note:
The CareSens PRO Control Solution can be purchased
separately. Please contact your authorised i-SENS sales
representative.