Step 10
This setting allows you to turn the hypoglycemia indicator (possible
low blood sugar) on or off and to select the desired level for the
indicator. You will be alerted any time your test result is lower than
the selected level. On pressing the
◀
or
▶
button, the screen
will display 'On' or 'OFF'. Press the
S
button when 'On' appears
to enter the setting. Press the
◀
or
▶
button until the desired
hypoglycemia level between
1
0 and 90mg/dL (
0.56
–5.0 mmol/L)
appears. Press the
S
button to confirm the hypoglycemia level and
go to the next step.
Setting the Hypoglycemia (HYPo) Indicator
Caution:
Ask your healthcare professional to help you decide what your
hypoglycemia level is before setting your level.
Note:
If the test result is lower than
the pre-set hypoglycemia
level, the meter will display
the following.
HYPo warning
will blink at the
bottom.
Step 9
This setting allows you to turn the strip expiration date indicator
on or off. This setting turns the function on or off only. See page
20 to set the strip expiration date.
When 'EP' appears on the screen, press the
◀
or
▶
button. The
screen will display 'On' or 'OFF'. Press the
S
button to confirm
the setting.
If you do not want to set the indicator, press the
S
button while
the screen displays 'OFF'.
Turning on the Strip Expiration Date Indicator
Note:
If the pre-set expiration date
is reached, the meter will
display 'EP' when the test strip
is inserted and when the test
result is displayed right after
the test.
If the expiration date is set to
October of 20
20
, the meter will
display 'EP' at the beginning of
November, 2020.