background image

101-3GM782-011

EN

Rev

. Dat

e: 2017-06

BIONIME GmbH

Tramstrasse 16, 9442 Berneck

Switzerland

E-mail: [email protected]

0197

BIONIME CORPORATION

No. 100, Sec. 2, Daqing St., South Dist.,

Taichung City 40242, Taiwan

Tel:   +886 4 2369 2388

Fax:  +886 4 2261 7586

http: //www.bionime.com

E-mail: [email protected]

USER’S MANUAL

Blood Glucose Monitoring System

GM700SB

Содержание GM700SB

Страница 1: ...Berneck Switzerland E mail info bionime ch 0197 BIONIME CORPORATION No 100 Sec 2 Daqing St South Dist Taichung City 40242 Taiwan Tel 886 4 2369 2388 Fax 886 4 2261 7586 http www bionime com E mail inf...

Страница 2: ......

Страница 3: ...d by Bionime Corporation If you have any questions or concerns please contact your local Bionime Customer Service or email to rightest bionime com TM Thank you for selecting the Rightest Blood Glucose...

Страница 4: ...Test Strips are exposed to a substantial change in temperature please wait 30 minutes before measurement Follow all environmental protection regulations when disposing of batteries You may manually t...

Страница 5: ...59 38 39 42 44 45 46 About the Blood Glucose Monitoring System The Blood Glucose Monitoring System The Blood Glucose Meter The Blood Glucose Test Strip Before Testing Meter Activation and Battery Cha...

Страница 6: ...032 battery installed 6 Blood Glucose Test strips GS700 10 25 pcs 7 Control Solution GC700 Normal Level is provided in your starter kit 8 Lancing Device 9 Clear Cap 10 Disposable Sterile Lancets 10 pc...

Страница 7: ......

Страница 8: ......

Страница 9: ......

Страница 10: ...Rightest replacing the battery all symbols will appear on the screen 6 You must set the time and date when replacing the battery See Chapter Setting Up Your Meter Setting the Date Time and Volume on p...

Страница 11: ...tery installation steps to replace battery The Rightest Blood Glucose Meter will perform a self test Press the main button to exit the self test and enter the Setting Mode When the meter is ON hold do...

Страница 12: ...r Set up for over 2 minutes the Blood Glucose Meter will leave setting mode and power off automatically Meter default is set according to your local preference TM Rightest GM700SB NOTE The date and ti...

Страница 13: ...2 3 1...

Страница 14: ...u have to perform meal marker setting before testing 1 Before test press the Left or Right button to choose between the before meal and after meal then press main button to confirm and do the test aft...

Страница 15: ......

Страница 16: ...eep if volume is turned on Apply the blood sample within 2 minutes 13 Place the lancing device against your fingertip and press the release button 14 Touch and hold the blood drop to the edge of sampl...

Страница 17: ...ne including other family members All parts of this kit are considered biohazards and can potentially transmit infectious diseases even after you have performed the cleaning and disinfecting procedure...

Страница 18: ...cted mobile device When the test result uploading has completed Bluetooth will disconnect from mobile App automatically Data uploading The GM700SB meter is now uploading test results to your mobile de...

Страница 19: ...A...

Страница 20: ...GS700 TM Use each Rightest Blood Glucose Test Strip GS700 immediately after removing it from the vial TM Do not reuse Rightest Blood Glucose Test Strips GS700 Do not perform testing immediately after...

Страница 21: ......

Страница 22: ......

Страница 23: ...ol Solution Range printed on the Blood Glucose Test Strip GS700 vial label TM Rightest TM Rightest CAUTION Your Control Solution Test results will not be included in the average calculations however t...

Страница 24: ...on on the side to review all previous results with date and time Results display from the most recent Sequence no 1 to the oldest Sequence no 500 in the lower right hand corner of the screen 3 The Qua...

Страница 25: ...test results and test results under abnormal temperature 6 C 43 F 44 C 111 F will be excluded from average calculations The Average function requires that the correct time and date is set 1 Press the...

Страница 26: ......

Страница 27: ......

Страница 28: ...l L Not for screening or diagnosis of diabetes mellitus Not for use on critically ill patients in shock dehydrated patients or hypo osmolar patients Alternative site testing AST should only be perform...

Страница 29: ......

Страница 30: ...s to our customers Please review all the instructions to make sure you are performing the steps correctly If you have any questions or in case of problems with the products please contact your local B...

Страница 31: ...nd Disposable Sterile Lancets Manufacturer SteriLance Medical SuZhou Inc No 68 LiTangHe RD XiangCheng SuZhou JiangSu 215133 P R China Product complies with Medical Device Directive 93 42 EEC CE 0197 E...

Отзывы: