48 www.i-sens.com
www.i-sens.com 49
Difference Averages
No.
Interferent
17
L-Dopa
0.04 mmol/L
0.5%
18
Maltose
-0.4 mmol/L
-1.1%
19
Mannitol
0.1 mmol/L
-0.7%
20
Methyldopa
-0.03 mmol/L
0.2%
21
Pralidoxime Iodide
0.0 mmol/L
1.4%
22
Salicylate
0.1 mmol/L
-0.1%
23
Tolazamide
-0.3 mmol/L
-2.8%
24
Tolbutamide
-0.3 mmol/L
-7.3%
25
Triglycerides
-0.1 mmol/L
4.7%
26
Uric acid
-0.1 mmol/L
0.5%
27
Xylose
-0.04 mmol/L
-1.0%
Interval1
Interval 2
(2.8–5.5 mmol/L) (13.9–19.4 mmol/L)
User Performance Evaluation
A study evaluating glucose values from fingertip capillary blood
samples obtained by 100 lay persons showed the following
results:
100% within ± 0.83 mmol/L of the medical laboratory values
at glucose concentrations below 5.55 mmol/L and 98.3%
within ±15% of the medical laboratory values at glucose
concentrations at or above 5.55 mmol/L.
Warranty Information
Manufacturer’s Warranty
i-SENS, Inc. warrants that the CareSens N POP Meter shall be
free of defects in material and workmanship in normal use for
a period of five (5) years. The meter must have been subjected
to normal use. The warranty does not cover improper
handling, tampering, use, or service of the meter. Any claim
must be made within the warranty period.
i-SENS will, at its discretion, repair or replace a defective meter
or meter part that is covered by this warranty. As a matter
of warranty policy, i-SENS will not reimburse the consumer’s
purchase price.
Obtaining Warranty Service
To obtain warranty service, you must return the defective
meter or meter part along with proof of purchase to your
nearest i-SENS sales or customer service representative.
Содержание CareSens N Pop