![i-SENS CareSens N Pop Скачать руководство пользователя страница 4](http://html1.mh-extra.com/html/i-sens/caresens-n-pop/caresens-n-pop_owners-booklet_602498004.webp)
4 www.i-sens.com
www.i-sens.com 5
Important Information: Read This First!
For optimum safety and benefits, please read the entire manual
contents before using the system.
Intended use:
CareSens N POP Blood Glucose Monitoring System is used for
the quantitative measurement of the glucose level in capillary
whole blood as an aid in monitoring diabetes management
effectively at home or in clinical settings. CareSens N POP
Blood Glucose Monitoring System should be used only for
self-testing outside the body (
in vitro
diagnostic use only).
CareSens N POP Blood Glucose Monitoring System should not
be used for the diagnosis of diabetes or for testing newborns.
Testing sites include the traditional fingertip testing along
with alternate sites testing on forearm and palm.
• Glucose in blood samples reacts with the chemical in the
test strip to produce a small electrical current. The CareSens
N POP meter detects this electrical current and measures
the amount of glucose in the blood sample.
• The CareSens N POP Blood Glucose Meter is designed to
minimise code related errors in monitoring by using the
no-coding function.
• The CareSens N POP Blood Glucose Meter should be used
only with the CareSens N test strips.
• An abnormally high or low red blood cell count (hematocrit
level over 65% or below 15%) may produce inaccurate
results.
• If your test result is below 3.3 mmol/L or above 13.3 mmol/L,
consult a healthcare professional immediately.
• Inaccurate results may occur in severely hypotensive
individuals or patients in shock. Inaccurate low results
may occur for individuals experiencing a hyperglycemic-
hyperosmolar state, with or without ketosis. Critically ill
patients should not be tested with blood glucose meters.
•
Inaccurate results may occur in patients undergoing
oxygen therapy.
If you need assistance, please contact your authorised i-SENS
sales representative or visit www.i-sens.com for more information.
Important Information
Meaning of Symbols Used:
For
in vitro
diagnostic use
This product fulfills the requirements for Directive 98/79/EC
on
in vitro
diagnostic medical devices
Cautions for safety and optimum product use
Use by (unopened or opened test strip vial)
Do not discard this product with other household-type waste
Do not reuse
Batch code
Consult instruction for use
Manufacturer
Temperature limitation
Serial number
Authorised representative
Содержание CareSens N Pop