30 www.i-sens.com
www.i-sens.com 31
Step 1
Unscrew the lancing device tip.
Step 2
Stick the lancet into the saved
protective disk.
Push the lancet ejector forward with
the thumb to dispose of the used lancet
in a proper biohazard container.
Discarding Used Lancets
Caution:
The lancet is for single use only. Never share or reuse
a lancet. Always dispose of lancets properly.
Step 11
Apply the blood sample to the narrow
end of the test strip until you hear a
beep. At this time, the display segments
will rotate clockwise while the blood is
going in.
Test result will appear after the meter
counts down from 5 to 1. The result will
be automatically stored in the meter’s
memory.
If the test strip is removed after the test
result is displayed, the meter will automatically switch off after
3 seconds.
Discard used test strips safely in disposable containers.
Note:
The meter may switch off if the blood sample is not
applied within 2 minutes of the symbol appearing on the
screen. If the meter turns off, remove the strip and reinsert it,
and start from Step 7.
Caution:
Do not allow any foreign substances, such as dirt, blood, or
water, enter into the meter. The meter may be damaged or
may malfunction. Follow the warning information provided
below to prevent possible damage to the meter.
• Do not apply the blood sample directly to the test strip port.
• Do not apply the blood sample to the test strip while
holding the meter in a way that the tip of the test strip faces
upwards. The blood sample may run down the surface of
the test strip and flow into the test strip port.
• Do not store your meter in unsanitary or contaminated sites.
Содержание CareSens N Pop
Страница 27: ...50 www i sens com MEMO ...