Understanding Error Messages
Message
What It Means
A used test strip was
inserted.
The blood sample has
abnormally high
viscosity or
insufficient volume.
There is a problem
with the meter.
An electronic error
occurred during the
test.
This error message
may appear when the
wrong blood glucose
test strip is used
instead of the
CareSens N test strip.
What To Do
Repeat the test with a
new test strip.
Repeat the test with a
new test strip.
Do not use the meter.
Contact your
authorised i-SENS sales
representative.
Repeat the test with a
new test strip. If the
error message persists,
contact your
authorised i-SENS sales
representative.
Repeat the test with a
CareSens N test strip.
The blood or control
solution sample was
applied before the
symbol appeared.
Repeat the test with a
new test strip and wait
until the symbol
appears before
applying the blood or
control solution
sample.
Message
What It Means
The temperature
during the test was
above the operating
range.
What To Do
Note:
If the error messages persist, contact your authorised
i-SENS sales representative.
The temperature
during the test was
below the operating
range.
Move to an area where
the temperature is
within the operating
range (5-50˚C/41-
122°F) and repeat the
test after the meter
and test strips have
reached a temperature
within the operating
range.