Firmware Version 2.0
MultiSystem 4070
110
Operating software
ENG
Trigger mode
Define the trigger:
•
INACTIVE
The trigger is not enabled
•
CHANNEL
Definition of a measurement channel as trigger
•
KEY
Triggering at the press of a key
•
TIMER
Definition of a trigger time
Definition of a measurement channel as trigger
1
For the
Trigger mode
, select the
CHANNEL
option.
2
Open the
Trigger condition
dialog entry and highlight the desired option.
Then press
.
•
GREATER
: Actuation when trigger value is exceeded
•
LOWER
: Actuation when trigger value has fallen below
•
RISING
: Actuation when trigger value has fallen below by more than
5% and then exceeded, “rising edge”
•
FALLING
: Actuation when trigger value is exceeded by more than 5%
and has then fallen below, “falling edge”
3
Enter trigger value
.
4
Confirm changes and exit the dialog:
Definition of a trigger time
1
For the
Trigger mode
, select the
TIMER
option.
2
Enter the date of the trigger time –
.
3
Enter the time of day of the trigger time –
.
4
Confirm changes and exit the dialog: