Error messages / troubleshooting
FAM 15/30/50/70 ATEX
en(us)
Page 51 / 96
BeWa FAM15-70 Atex 3193049h en(us) 2013-06-19
2013-06-19
[H6] Inlet (IN) clogged
Step
Description
1.
Malfunction:
"Inlet clogged"
2.
Cause: No medium is making it into the vacuum column
3.
Check the setting of the shut off fittings in the suction line. If
they are closed, then open them. Check/clean the suction
screen.
4.
Press the RESET button to acknowledge/reset the fault.
5.
Does the fault repeat after ~ 4 minutes?
-> a. Yes
-> proceed to Step 7.
6.
-> b. No
-> proceed to Step 8.
7.
Check the following points:
-
Check the suction pipe for cross-section constriction.
-
Check the functioning of the vacuum pump in manual mode.
-
Are all locking features in the suction line open?
8.
The failure is repaired.