Error messages / troubleshooting
FAM 15/30/50/70 ATEX
en(us)
Page 50 / 96
BeWa FAM15-70 Atex 3193049h en(us) 2013-06-19
2013-06-19
[H5] If the vacuum column is overfilled,
Step
Description
Malfunction:
"Vacuum column overfilled"
1.
Press the RESET button.
2.
Select manual menu 1, and start the evacuation pump
3.
Can the vacuum column be emptied within 3.5°minutes from
level 03 to level 01?
-> a. Yes
-> proceed to Step 5.
4.
-> b. No
-> proceed to Step 6.
5.
Excessive foaming -> Vacuum in the vacuum column too high -
> Reduce the vacuum (e.g. from 300 mbar
(abs)
-> 400 mbar
(abs)
)
6.
Check the evacuation pump for proper functioning.
7.
The fault is repaired.