Performing maintenance
CTU 2xxx series
en-US
Page 83 / 108
BeWa CTU2000 3160738h en-us 2018-11-15.docx
2018-11-15
Reservoir filling/refilling
To fill/refill the reservoirs with test fluid, follow the instructions given on page
50.
Checking the gloves
The gloves are subject to wear both mechanically and chemically.
Mechanical wear - cuts and tears result from components with sharp edges
and wear can be accelerated by components with rough surfaces.
For details of wear due to chemical action, for example by permeation,
please see page 84.
Provided there is no mechanical damage to the surface, the gloves can be
used for six months (with average use & single shift operation).
The tests that we carried out were based on the "G 60 Spezial" analysis fluid.
The service life depends on the nature of the analysis fluid used and can
vary.
CAUTION
Worn gloves
Health hazard
►
Check the gloves for leaks on a daily basis
►
If you find signs of wear, replace the gloves
immediately
►
Replace the gloves every six months