Operating the text display TD 200
CTU 2xxx series
en-US
Page 48 / 108
BeWa CTU2000 3160738h en-us 2018-11-15.docx
2018-11-15
Selecting the operating mode
The operating mode is set using the selector switch at the front.
Stand-by
When you switch the unit to stand-by:
•
The compressed air line is blocked off.
•
The reservoirs are adjusted to normal pressure and then sealed off.
•
The fan power is reduced.
•
The lighting is switched off.
The clean room remains operational in stand-by mode.
Compressed air continues to flow via the vacuum generator.
This does not represent a system defect!
Operation
When you switch the unit to operation:
•
The compressed air line is opened.
•
The reservoirs are adjusted to operating pressure.
•
The fan power is increased to operating power.
•
The lighting is switched on.