Opening/closing the door
CTU 2xxx series
en-US
Page 35 / 108
BeWa CTU2000 3160738h en-us 2018-11-15.docx
2018-11-15
Installing gloves
Prepare the gloves before installation. To do so, roll up the gloves from the back for
about 5 cm to form a roll, like in the following illustration.
Make sure that the text / labelling on the inside of the glove is on
the side into which you put your hand.
The text / labeling can come undone during sampling and lead to
impairment of the analyses.
Introduce the gloves through the openings into the analysis chamber. Then, stretch
the rolled end of the glove over the access opening from the outside. It is not
necessary to additionally secure the gloves to the access opening.
Use electro-statically conductive gloves
When replacing the gloves, ensure that there is sufficient protection (i.e. conductivity)
from electrostatic charge as specified in IDN EN 61340-5-1.
R
g
= 2.0 x 10
8
Ω
Please make sure that resistance R
g
(
Ω
) is between 7.5x10
5
≤ R
g
≤ 1x10
12
to EPA
ground or a grounding point for gloves and finger stalls.