94
6.6.3.
Replacement of consumables and fuse
1.
Replacing chinrest paper
①
Pull out two fixing pins from chinrest.
②
Put a new chinrest paper on the chinrest.
③
Insert two fixing pins into the chinrest paper hole.
④
Attach the chinrest paper to the chinrest.
2.
Replacing fuse
①
Ensure the power switch of device off (O).
②
Remove power cable from inlet.
③
Pull out fuse holder in the inlet with a tweezers.
④
Replace two new fuses in the fuse holder. Be sure to check the fuse specification for the
replacement (250V T 3.15AL).
⑤
Insert fuse holder into the inlet.
6.6.4.
Calibration
Note: Huvitz recommends to calibrate the system once a year.
Contact to the HUVITZ’s service technicians or other authorized experts