42
satisfactory though the SSI is low, you can ignore SSI and save the result.
④
If the image is satisfactory, press OK icon (
) to save image.
⑤
If the image is not satisfactory, press retry icon (
) and retry image capturing.
A.
If fundus image result too bright or too dark because of lighting, regulate the flash
intensity using flash icon(
) in observation
mode.
B.
If fundus image is too dark because of small pupil size of patient, try small pupil
mode by using small pupil icon (
) in observation mode.
C.
Try moving internal fixation target position by pressing fixation icon (
)
and changing position of green cross (
) if needed.
When green cross position changes, the position of internal fixation target is also
changed.
D.
Try to change scan position by dragging scan range while scan range icon turned
on. If reset scan position icon(
) is pressed, scan position moves to the
default center position.
10.
Capture image and check image quality (in OCT mode)
10-1. When alignment and focus is done, press scan start icon (
) to start OCT scan.
If alignment and focus is in good condition, scan start function is accomplished automatically, and
scan start icon changes to optimize icon(
)
10-2. Optimize OCT signal by pressing optimize icon on the screen or optimize button on the body (A).
10-3. Adjust the vertical position of retina image in the screen by moving reference mirror using slide
bar. (
).
10-4. Press the button on joystick to capture image.
If auto shooting is on, ‘10-2. Optimize’ and ‘10.4 Capture’ is accomplished automatically.
Содержание HOCT-1F/1
Страница 1: ...OPTICAL COHERENCE TOMOGRAPHY HOCT 1F 1 USER MANUAL ...
Страница 4: ...4 9 EMC INFORMATION 103 10 SERVICE INFORMATION 105 11 SOFTWARE LICENSE AGREEMENTS 106 ...
Страница 18: ...18 Rear View ...
Страница 62: ...62 1 19 Selecting FULL Screen Icon shows the current Bscan image in full screen ...
Страница 73: ...73 HOCT 1F 1 1 19 Selecting FULL Screen Icon shows the current Bscan image in full screen ...
Страница 102: ...102 8 3 Drawings of System ...