![Huvitz HOCT-1F/1 Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/huvitz/hoct-1f-1/hoct-1f-1_user-manual_2190284048.webp)
48
⑥
If the image is not satisfactory, press retry icon (
) and retry image capturing.
A.
If fundus image result too bright or too dark because of lighting, regulate the flash
intensity using flash icon(
) in observation
mode.
B.
Try moving internal fixation target position by pressing fixation icon (
)
and changing position of green cross (
) if needed.
When green cross position changes, the position of internal fixation target is also
changed.
C.
Try to change scan position by dragging scan range while scan range icon turned
on. If reset scan position icon(
) is pressed, scan position moves to the
default center position.
4.
Capture anterior chamber angle image and check image quality (in ACA line mode)
4-1. Alignment and focus
①
Guide patient to gaze left-hand side or right-hand side according to the interested part.
②
Move body with joystick slowly to align anterior scan line (A) and objective area of
capturing (B).
③
Start OCT scan by pressing scan start icon (
).
④
Move body with joystick slowly until anterior chamber appears on the screen.
4-2. Measuring anterior chamber angle
①
Optimize OCT signal by pressing optimize icon on the screen (
) or optimize button
on the body (A).
Содержание HOCT-1F/1
Страница 1: ...OPTICAL COHERENCE TOMOGRAPHY HOCT 1F 1 USER MANUAL ...
Страница 4: ...4 9 EMC INFORMATION 103 10 SERVICE INFORMATION 105 11 SOFTWARE LICENSE AGREEMENTS 106 ...
Страница 18: ...18 Rear View ...
Страница 62: ...62 1 19 Selecting FULL Screen Icon shows the current Bscan image in full screen ...
Страница 73: ...73 HOCT 1F 1 1 19 Selecting FULL Screen Icon shows the current Bscan image in full screen ...
Страница 102: ...102 8 3 Drawings of System ...