![Huvitz HNT-1 / 1P Скачать руководство пользователя страница 47](http://html.mh-extra.com/html/huvitz/hnt-1-1p/hnt-1-1p_service-manual_2190281047.webp)
46 Non-Contact Tonometer HNT-1/1P ------------------------------------------------------------------------------------------------
i.
Index: The sequence of logs written
ii.
Cat: Category. This number shows which operation mode when the
log written.
iii.
Time: Time with format YY/MM/DD hh:mm:ss.
iv.
Func#: The function number. This number indicates specific internal
function code.
v.
Ret: The return value of the function. This shows the result of the
function.
vi.
Page: Current page/total number of page.
vii.
Checker: The checker number shows the model.
1.
0xB3: HNT-1P
2.
0xB2: HNT-1
viii.
nxt wr idx: Next write index.
ix.
Cycle cnt: The log space is limited in the physical memory. So the
number of log index has a cycle of writing in that space.
x.
Func detail: The internal function code name.
xi.
Res meaning: OK or Failed as the result of the function.
(3)
Recent: To move to the current index log.
(4)
PgUp: To move above page.
(5)
PgDn: To move below page.
(6)
IndexUp: To move above index.
(7)
IndexDn: To move below index.
4-1-14. Print Engineer Settings
[Purpose
]
Print current engineer setting value.
[Execution]
(1)
Select the
“Print Engineer Settings” button at Engineer Mode.
(2)
The following items are displayed in the print.
①
DPC
: dead pixel collection
②
SOL CTRL
: solenoid control setting value
③
FOCUS LED SETUP
: focus led setting value