40 Non-Contact Tonometer HNT-1/1P ------------------------------------------------------------------------------------------------
(5)
Press the “Start” button to automatically measure the target.
(6)
Then it will start automatically to measure 5 times and display 5 results.
(7)
Finally, the result of the CCT calibration is output at the bottom.
(8)
If you press the “Check” button, You can check that the calibrated CCT value is
working properly.
(9)
To calibrate the R,
press the table space to start “R Setup”. You can set the
target
R using the “+”, “-“, “++”, “--“ buttons.
(10)
Press the “Start” button to automatically measure the target.
(11)
Then it will start automatically to measure 5 times and display 5 results.
(12)
Finally, the result of the R calibration is output at the bottom.
(13)
If you press the “Check” button, You can check if the calibrated R value is
working properly.
(14)
If you want to see the front view, press the
“<- ->” button.