42 Non-Contact Tonometer HNT-1/1P ------------------------------------------------------------------------------------------------
as pen. When finish touching one cross shape, the next cross shapes are
displayed.
(4)
When done, press
“OK” button and the result values(calibrated position of X, Y)
will be output.
4-1-10. Focus Align Test
[Purpose
]
Test the focus LED, track motor setup and track dist setup.
[Test Procedure]
(1)
Enter the Focus Align Test at Engineer Mode.
(2)
Press
“Start” button on the screen to check the auto tracking & focusing
movement.
(3)
Check the results measured at the bottom and right side of the screen.
LUSP
: Left Up Spot Point of X, Y
LUSZ
: Left Up Spot Size
LUOF
: Left Up Offset
LDSP
: Left Down Spot Point of X, Y
LDSZ
: Left Down Spot Size
LDOF
: Left Down Offset
ASP
: Aiming Spot Point of X, Y
ASZ
: Aiming Spot Size
AOF
: Aiming Offset