38 Non-Contact Tonometer HNT-1/1P ------------------------------------------------------------------------------------------------
(3)
For manual mode, change the following items.
A.
MA
– TARGET DIST(X)
: x direction
B.
MA
– TARGET DIST(Y)
: y direction
C.
MA
– TARGET DIST(ZF)
: front side on z direction
D.
MA
– TARGET DIST(ZR) : rear side on z direction
(4)
For Auto Focus mode, change the following items.
A.
AF
– TARGET DIST(X)
: x direction
B.
AF
– TARGET DIST(Y)
: y direction
C.
AF
– TARGET DIST(ZF)
: front side on z direction
D.
AF
– TARGET DIST(ZR)
: rear side on z direction
(5)
Press
“Clear” button to reset all changes to default settings.
(6)
Press
“Save” button to save the current settings.
4-1-8. Pachy Measure Setup(only HNT-1P)
[Purpose]
Setup pachymetry
[Preparation]
(1)
Enter into Pachy Measure Setup Mode at Engineer Mode.
(2)
To increase the accuracy of pachy measurements, you can set several
parameters according to your environment.