18
Setup
2. Dock TOCO-transducer for charging again:
•
charge bay LED and battery indicator LEDs turn on
3. Remove US-transducer (US):
•
charge bay LED turns off
•
LED in US-transducer starts fl ashing
•
US-transmission indicator LED turns green
•
US battery indicator LEDs indicate charge status
•
US channel becomes active: US is indicated on fetal monitor
4. Simulate audio signal:
•
Stroke bottom face of transducer at approximately 2 strokes per second to
simulate a fetal heart signal.
•
fetal monitor shows heart rate, after a short delay.
5. Remove the TOCO-transducer:
•
charge bay LED turns off
•
LED in TOCO-transducer starts fl ashing
•
TOCO-transmission indicator LED turns green
•
TOCO battery indicator LED's indicate charge status
•
TOCO channel becomes active: TOCO is indicated on fetal monitor
Note :
The TOCO transducer must be in operating range of the US
transducer (Maximum range 30cm).
6. Press gently on the sensor area of the transducer:
•
fetal monitor shows adequate TOCO values
7. Dock US-transducer for charging
•
US and TOCO-transmission indicator LEDs turn Amber
•
charge bay LED turns on
•
US battery bay LEDs indicate charge status
•
Fetal monitor returns to normal operating mode with wired transducers.
8. Dock TOCO transducer for charging
•
charge bay LED turns on
•
TOCO battery indicator LEDs indicate charge status
9. Wait until both Battery Indicators show all 4 LEDs on (battery fully
charged). A complete charge cycle takes approximately 2.5 hours.
Note :
It is recommended that the serial numbers from installed and
matched receiver and transducer units are recorded for future
use if required. This may be important if several telemetry
systems are used within a ward or confi ned area.