No TOCO signal on fetal monitor from telemetry.
POSSIBLE CAUSE
SOLUTION
US-transducer not receiving
signal from TOCO.
Bring the transducers closer together
until US receives signal from TOCO.
TOCO not running (LED not
fl ashing).
Charge TOCO-transducer and try again.
TOCO baseline zeroed before
wireless system has stabilised
Wait for 20-30s after activating the
wireless system before using the fetal
monitor contractions zero function. Note
that it may be necessary to repeat the
contractions zero function after any
drop-out in the wireless link, allowing the
system to re-stabilise on each occasion.
Charging panel or contacts show corrosion.
POSSIBLE CAUSE
SOLUTION
Transducer was wet or
contaminated with gel when
docked on the receiver.
Always clean and dry the transducer
before docking and charging. If
necessary have the contact panel
replaced.
Receiver or Transducer does not respond
POSSIBLE CAUSE
SOLUTION
The discharge of static
electricity may have disabled the
receiver.
Disconnect the power cord for about 5
seconds. Re-connect the cable.
The discharge of static
electricity may have disabled the
transducer.
Refer to section 8 for re-setting
transducers.
Green LED of undocked Transducer does not have normal operational pattern.
POSSIBLE CAUSE
SOLUTION
No LED (Not paired- Normal
operation)
No LED
Unusual green LED pattern
Solid green LED
Transducers may be locked
Pair transducers (See Section 4)
Redock transducer, wait for identifi cation.
Undock and recheck. Try 3 times.
Manually Switch off Transducers (Refer
to 8.1 and 8.2)
If problem persists, transducer is faulty.