e and Cleaning
6. Care and Cleaning
6.1 General
Care
Although the Freedom is robust and has been designed to withstand normal
clinical use, the unit contains delicate components, which should be handled
and treated with care.
Periodically, and whenever the integrity of the system is in doubt, carry out a
check of all functions as described in Section 4.2. If there are any defects to
the housing of a transducer do not use it in water. Contact Huntleigh or your
distributor to get it repaired or to order a replacement.
If a transducer is dropped, check that the seal and the housing are not
damaged. If you are in doubt please contact Huntleigh or your distributor for
further instructions.
Please ensure that you check with your facility’s local infection
control policy and medical equipment cleaning procedures.
Observe warnings and guidance on cleaning fl uid labelling
regarding use and personal protective equipment (PPE).
Do not use abrasive cloths or cleaners on the transducer,
receiver or accessories.
Do not use automatic washers or autoclaves to clean the
transducers.
Phenolic detergent based disinfectants, solutions containing
cationic surfactants, ammonia based compounds or perfumes
and antiseptic solutions such as Steriscol or Hibiscrub should
never be used on any part of the Freedom system.
If detergent or disinfectant wipes are used to clean the
transducers or the receiver ensure that excess solution is
squeezed from the wipe prior to use.
Always switch off the Receiver by disconnecting the unit from
the AC supply before cleaning and disinfecting, do not allow any
fl uid to enter the Receiver and do not immerse the receiver in
any solution.