80
The following table is intended to help the user if they encounter any
problems:-
25. Fault Finding
F
ault Finding
Symptom
Possible cause/remedy
Green power indicator
not illuminated
Power cable not connected to live power
source
Defective power cable
Mains input fuses blown
Amber unit ‘ON’.
Indicator not illuminated
Unit is not switched on
No signal on trace
Defective patient cable
Excessively noisy trace
Electrode site not properly prepared
Poor electrode contact
Defective patient cable
No heart rate display or
flashing heart symbol
Patient electrodes incorrectly sited
No QRS bleep
Volume set too low
Continuous alarm
indications. No ECG
signal on display (LEAD
OFF displayed)
Defective patient cable Electrode or lead off
Faulty cuff connection
and related NIBP
errors indicated on
screen display
See table in NIBP section
No temperature digits
Check connection of probe to module
Replace temperature sensor
Temperature digits
reading 20 degrees
Replace temperature sensor