50
Invasive Blood P
ressure Monitoring
System calibration is automatic, however the user
can check the system by pressing the
CAL button on
the transducer.
The appropriate IBP channel will display 100 on the
screen.
17.4 Calibrating the System
Precautions: -
Avoid conductive connections with other monitoring devices.
Allow the system to warm up for 5 minutes before calibrating.
Do not use a syringe smaller than 5 ml.
Do not use excessive force on the syringe while attaching fluid-filled
components.
Do not twist cables.
Do not autoclave or ethylene oxide sterilise transducers, interface
cables or clamps.
If any part of the equipment becomes accidentally wetted during
use, disconnect from the patient, isolate from the mains supply and
dry the parts before reconnecting.
100mmHg
17.3 Zeroing the system
Each channel can be set up / calibrated independently.
Allow the transducer to warm up for 5 minutes before zeroing.
Vent the transducer to atmospheric pressure by turning the ZERO
stopcock off to the patient.
Touch the appropriate IBP area of the screen to display the set-up
screen and press ZERO to zero the transducer.
A two second delay occurs while the zero adjustment takes place.